Arca
Skip to content

Itst Intertrochanteric Subtrochanteric Intramedullary Femoral Nail

Rod, Fixation, Intramedullary And Accessories. 510(k) K032367, from Zimmer, Inc. (Warsaw, IN). Product code HSB, class 2.

As of Nov 17, 2026, K032367 is an FDA 510(k) held by Zimmer, Inc., substantially equivalent on Aug 12, 2003 after 11 days, with 12 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Aug 12, 2003
  • Review time: 11 days
  • Recalls: 12
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The ITST Intramedullary Nail is indicated for use in a variety of femoral fractures, such as: • Subtrochanteric Fractures • Intertrochanteric Fractures • Comminuted Fractures • Segmental Fractures • Fractures with Bone Loss • Proximal and Distal Fractures • Nonunions

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K032367, newest first; each opens FDA's recall record
Feb 28, 2017Z-2565-2017Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.Terminated
Feb 28, 2017Z-2562-2017Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.Terminated
Feb 28, 2017Z-2563-2017Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.Terminated
Feb 28, 2017Z-2564-2017Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.Terminated
Jan 11, 2016Z-1153-2016As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.Terminated
Jan 11, 2016Z-1138-2016As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.Terminated
Jan 11, 2016Z-1112-2016As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.Terminated
Oct 11, 2013Z-1014-2014Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations…Terminated
Oct 11, 2013Z-1012-2014Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations…Terminated
Oct 11, 2013Z-1013-2014Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations…Terminated
Jun 5, 2009Z-1702-2009Reamers labeled as having a standard diameter may actually be the dimensions of an Asia reamer, and vice versa.Terminated
Jun 5, 2009Z-1701-2009Reamers labeled as having a standard diameter may actually be the dimensions of an Asia reamer, and vice versa.Terminated

Other HSB Clearances

The newest 510(k)s for the same product code; their median review was 192 days; see every HSB device

Record

Type
Special
Received
Aug 1, 2003
Review panel
Orthopedic
Regulation
21 CFR 888.3020

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.