I-Stat 200 Portable Clinical Analyzer

Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph. 510(k) K940918, from I-Stat Corp. (Princeton, NJ). Product code CHL, class 2.

As of Nov 17, 2026, K940918 is an FDA 510(k) held by I-Stat Corp., substantially equivalent on May 10, 1994 after 74 days, with 19 recalls tied to it in FDA's recall database.

  • Substantially equivalent: May 10, 1994
  • Review time: 74 days
  • Recalls: 19
  • Cited as a predicate by: 2 devices

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K940918, newest first; each opens FDA's recall record
Apr 1, 2026Z-2000-2026Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than…Open, Classified
Apr 1, 2026Z-1999-2026Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than…Open, Classified
Aug 21, 2025Z-2582-2025Lack of a 510(k) premarket clearance for the i-STAT CG4+ cartridge to reflect updated sample type for lactate and measurement ranges for pH and PO2.Open, Classified
Dec 22, 2020Z-0881-2021Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100…Terminated
Dec 22, 2020Z-0880-2021Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100…Terminated
Sep 13, 2018Z-0615-2019Inaccurate potassium results may be generated. When testing with control fluid and/or patient samples with low or reference range potassium concentration, elevated potassium results may be generated. When testing with…Terminated
Jan 1, 2014Z-1306-2014Abbott Point of Care has determined that some individual patient results generated with the i-STAT G3+ cartridge lot N13183 have the potential to exhibit incorrectly elevated PCO2 and depressed pH results.Terminated
Jan 1, 2011Z-2133-2012i-Stat EG7 cartridges may generate falsely depressed PCO2 results for samples above 26 mm hg when stored at room temperature for four weeks or longer.Terminated
Sep 20, 2010Z-0594-2011i-STAT Analyzers batteries may overheat due to a circuitry failure, resulting in the analyzer becoming uncomfortably hot in the area of the battery component.Terminated
Sep 20, 2010Z-0595-2011i-STAT Analyzers batteries may overheat due to a circuitry failure, resulting in the analyzer becoming uncomfortably hot in the area of the battery component.Terminated
Sep 20, 2010Z-0596-2011i-STAT Analyzers batteries may overheat due to a circuitry failure, resulting in the analyzer becoming uncomfortably hot in the area of the battery component.Terminated
Sep 20, 2010Z-0593-2011i-STAT Analyzers batteries may overheat due to a circuitry failure, resulting in the analyzer becoming uncomfortably hot in the area of the battery component.Terminated
Aug 3, 2010Z-0929-2011Two cartridge lots were identified that may generate falsely elevated PCO2 results.Terminated
Aug 3, 2010Z-0930-2011Two cartridge lots were identified that may generate falsely elevated PCO2 results.Terminated
Feb 5, 2010Z-1794-2010Abbott has determined that there is a dependence of the i-STAT Hct assay on the vertical pitch angle of the i-STAT Analyzer during the blood analysis cycle.Terminated
Feb 5, 2010Z-1793-2010Abbott has determined that there is a dependence of the i-STAT Hct assay on the vertical pitch angle of the i-STAT Analyzer during the blood analysis cycle.Terminated
Feb 23, 2009Z-1540-2009Equipment problem: i-STAT cartridges were not properly sealed as the packaging system jammed, causing compromised seals.Terminated
Feb 23, 2009Z-1541-2009Equipment problem: i-STAT cartridges were not properly sealed as the packaging system jammed, causing compromised seals.Terminated
Sep 25, 2006Z-1621-2011Certain lots of G3+ cartridges may be difficult to fill.Terminated

Other CHL Clearances

The newest 510(k)s for the same product code; their median review was 267 days; see every CHL device

Record

Type
Traditional
Received
Feb 25, 1994
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1120

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files