I-Stat 200 Portable Clinical Analyzer
Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph. 510(k) K940918, from I-Stat Corp. (Princeton, NJ). Product code CHL, class 2.
As of Nov 17, 2026, K940918 is an FDA 510(k) held by I-Stat Corp., substantially equivalent on May 10, 1994 after 74 days, with 19 recalls tied to it in FDA's recall database.
- Substantially equivalent: May 10, 1994
- Review time: 74 days
- Recalls: 19
- Cited as a predicate by: 2 devices
Cited as a Predicate
2 later 510(k) summaries name this device as a predicate- Avl Omni Modular AnalyzerK990092 · Avl Scientific Corp. · Jan 29, 1999
- Modification To Agilent Component Monitoring System (Cms), V24/V26 Patient Monitors,…K013624 · Philips Medizin Systeme Boblingen GmbH · Dec 3, 2001
Recalls
Recalls FDA ties to K940918, newest first; each opens FDA's recall record| Apr 1, 2026 | Z-2000-2026Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than… | Open, Classified |
| Apr 1, 2026 | Z-1999-2026Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than… | Open, Classified |
| Aug 21, 2025 | Z-2582-2025Lack of a 510(k) premarket clearance for the i-STAT CG4+ cartridge to reflect updated sample type for lactate and measurement ranges for pH and PO2. | Open, Classified |
| Dec 22, 2020 | Z-0881-2021Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100… | Terminated |
| Dec 22, 2020 | Z-0880-2021Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100… | Terminated |
| Sep 13, 2018 | Z-0615-2019Inaccurate potassium results may be generated. When testing with control fluid and/or patient samples with low or reference range potassium concentration, elevated potassium results may be generated. When testing with… | Terminated |
| Jan 1, 2014 | Z-1306-2014Abbott Point of Care has determined that some individual patient results generated with the i-STAT G3+ cartridge lot N13183 have the potential to exhibit incorrectly elevated PCO2 and depressed pH results. | Terminated |
| Jan 1, 2011 | Z-2133-2012i-Stat EG7 cartridges may generate falsely depressed PCO2 results for samples above 26 mm hg when stored at room temperature for four weeks or longer. | Terminated |
| Sep 20, 2010 | Z-0594-2011i-STAT Analyzers batteries may overheat due to a circuitry failure, resulting in the analyzer becoming uncomfortably hot in the area of the battery component. | Terminated |
| Sep 20, 2010 | Z-0595-2011i-STAT Analyzers batteries may overheat due to a circuitry failure, resulting in the analyzer becoming uncomfortably hot in the area of the battery component. | Terminated |
| Sep 20, 2010 | Z-0596-2011i-STAT Analyzers batteries may overheat due to a circuitry failure, resulting in the analyzer becoming uncomfortably hot in the area of the battery component. | Terminated |
| Sep 20, 2010 | Z-0593-2011i-STAT Analyzers batteries may overheat due to a circuitry failure, resulting in the analyzer becoming uncomfortably hot in the area of the battery component. | Terminated |
| Aug 3, 2010 | Z-0929-2011Two cartridge lots were identified that may generate falsely elevated PCO2 results. | Terminated |
| Aug 3, 2010 | Z-0930-2011Two cartridge lots were identified that may generate falsely elevated PCO2 results. | Terminated |
| Feb 5, 2010 | Z-1794-2010Abbott has determined that there is a dependence of the i-STAT Hct assay on the vertical pitch angle of the i-STAT Analyzer during the blood analysis cycle. | Terminated |
| Feb 5, 2010 | Z-1793-2010Abbott has determined that there is a dependence of the i-STAT Hct assay on the vertical pitch angle of the i-STAT Analyzer during the blood analysis cycle. | Terminated |
| Feb 23, 2009 | Z-1540-2009Equipment problem: i-STAT cartridges were not properly sealed as the packaging system jammed, causing compromised seals. | Terminated |
| Feb 23, 2009 | Z-1541-2009Equipment problem: i-STAT cartridges were not properly sealed as the packaging system jammed, causing compromised seals. | Terminated |
| Sep 25, 2006 | Z-1621-2011Certain lots of G3+ cartridges may be difficult to fill. | Terminated |
Other CHL Clearances
The newest 510(k)s for the same product code; their median review was 267 days; see every CHL device- ABL90 FLEX PLUS System, safeCLINITUBESK252207 · Radiometer Medicals Aps · Apr 10, 2026
- i-STAT CG4+ cartridge with the i-STAT 1 SystemK244014 · Abbott Point of Care, Inc. · May 2, 2025
- ABL90 FLEX PLUS SystemK240998 · Radiometer Medicals Aps · Dec 13, 2024
- i-STAT CG8+ cartridge with the i-STAT 1 SystemK230285 · Abbott Point of Care, Inc. · Oct 27, 2023
- Stat Profile Prime Plus Analyzer SystemK221900 · Nova Biomedical Corporation · Sep 29, 2023
- i-STAT G3+ cartridge with the i-STAT 1 SystemK223857 · Abbott Point of Care, Inc. · Sep 15, 2023
- GEM Premier 7000 with IQM3K223608 · Instrumentation Laboratory Company · Aug 10, 2023
- EasyStat 300K211559 · Medica Corporation · Jan 27, 2023
Record
- Type
- Traditional
- Received
- Feb 25, 1994
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1120
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files