Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer. 510(k) K936159, from Intermedics Orthopedics (Austin, TX). Product code JWH, class 2.

As of Nov 20, 2026, K936159 is an FDA 510(k) held by Intermedics Orthopedics, substantially equivalent on May 22, 1995 after 515 days, with 6 recalls tied to it in FDA's recall database.

  • Review time: 515 days
  • Recalls: 6
  • Cited as a predicate by: 7 devices

Cited as a predicate

7 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K936159, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 29, 2017Z-1033-2018The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.Terminated
Jan 11, 2016Z-0839-2016LDPE bag containing the implant adheres to the highly polished implant surface.Terminated
Feb 25, 2014Z-1249-2014, Z-1262-2014A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the…Terminated
Aug 30, 2013Z-0078-2014, Z-0075-2014The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.Terminated

Other JWH clearances

The newest 510(k)s for the same product code; their median review was 88 days; see every JWH device

Record

Type
Traditional
Received
Dec 23, 1993
Review panel
Orthopedic
Regulation
21 CFR 888.3560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files