Arca
Skip to content
Tools
Tools

Zimmer Persona Knee System

Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer. 510(k) K113369, from Zimmer, Inc. (Warsaw, IN). Product code JWH, class 2.

As of Nov 19, 2026, K113369 is an FDA 510(k) held by Zimmer, Inc., substantially equivalent on Mar 27, 2012 after 133 days, with 25 recalls tied to it in FDA's recall database.

  • Review time: 133 days
  • Recalls: 25+
  • Cited as a predicate by: 19 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

This device is indicated for patients with severe knee pain and disability due to: - Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. - Collagen disorders, and/or avascular necrosis of the femoral condyle. - Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

19 later 510(k) summaries name this device as a predicate; the 12 newest

Recalls

Recalls FDA ties to K113369, newest first; the 25 newest; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 7, 2017Z-1808-2018, Z-1807-2018, Z-1806-2018, Z-1805-2018There is a potential for intermittent cracks in the raw material batch used to produce the affected products. The cleanliness of the affected products could be compromised if cracks penetrate the surface of the…Terminated
Mar 17, 2016Z-1341-2016, Z-1340-2016The firm received a customer complaint establishing that the product (Persona PS Articular Surfaces)information shown on the box differed by sides. One side showed P/N: 42-5114-008-18/ Lot: 62632101 and the other side…Terminated
Jan 11, 2016Z-1157-2016As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.Terminated
Mar 10, 2015Z-1360-2015Zimmer, Inc is voluntarily recalling 22 lots of Persona Anterior Referencing Sizer with Locking Boom, part number 42-5099-088-10. Devices are being removed from distribution due to nonconforming to print specifications.Terminated
Dec 4, 2014Z-0853-2015The affected lot was supplied by contract manufacturer and are potentially subject to disassociation of the magnet cover. Zimmer received one complaint indicating the magnet cover of a Persona Cemented Tibial Broach…Terminated
Nov 25, 2014Z-1042-2015Potential for fracture of the tip of the product. The fractured tip is approximately 4mm x 7mm x 3.5mm in size. Retained tip may result in: pain, soft tissue irritation, soft tissue damage, or increased wear due to…Terminated
Oct 9, 2014Z-0198-2015Cleaning process validation failure.Terminated
Oct 7, 2014Z-0123-2015Affected product did not undergo heat treatment hardening step leading to the potential for deformation of the tip.Terminated
Aug 7, 2014Z-2580-2014, Z-2575-2014, Z-2578-2014, Z-2579-2014, Z-2576-2014, Z-2577-2014Reports of breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). The Persona primary knee surgical technique has been updated. This field action correction is to ensure that all distributors and…Terminated
Jun 10, 2013Z-2299-2014, Z-2297-2014, Z-2296-2014Zimmer had received complaints that reported breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). This is a retrospective recall which covers Zimmer's corrective actions taken earlier this year…Terminated
Nov 14, 2012Z-0548-2013, Z-0553-2013, Z-0554-2013, Z-0541-2013All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, and associated tools including the Tibial Drill Guide, Broach, and Sizing Plate are being removed and replaced with…Terminated

Other JWH clearances

The newest 510(k)s for the same product code; their median review was 88 days; see every JWH device

Record

Type
Traditional
Received
Nov 15, 2011
Review panel
Orthopedic
Regulation
21 CFR 888.3560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.