twiist system
Alternate Controller Enabled Insulin Infusion Pump. 510(k) K263756, from Deka Research and Development Corp. (Manchester, NH). Product code QFG, class 2.
As of Nov 28, 2026, K263756 is an FDA 510(k) held by Deka Research and Development Corp., substantially equivalent on Nov 12, 2026 after 27 days, with no recalls tied to it in FDA's recall database.
- Review time: 27 days
- Recalls: 0
- MiniMed Fit pumpK262992 · Medtronic MiniMed · Nov 13, 2026
- Tandem Mobi insulin pump with interoperable technologyK261816 · Tandem Diabetes Care, Inc. · Nov 12, 2026
- Mint ACE PumpK263422 · Beta Bionics, Inc. · Oct 21, 2026
- Mint ACE PumpK260001 · Beta Bionics, Inc. · Sep 11, 2026
- twiist Automated Insulin Delivery (AID) SystemK261672 · Deka Research and Development · Jun 17, 2026
- iLet ACE PumpK253976 · Beta Bionics, Inc. · Apr 30, 2026
- MiniMed Flex pumpK253743 · Medtronic Minimed, Inc. · Mar 12, 2026
- MiniMed 780G Insulin Pump (MMT-1884); MiniMed 780G Insulin Pump Kit (MMT-1894)K253470 · Medtronic Minimed, Inc. · Jan 15, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Oct 16, 2026
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 880.5730
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files