MiniMed Flex pump
Alternate Controller Enabled Insulin Infusion Pump. 510(k) K253743, from Medtronic Minimed, Inc. (Northridge, CA). Product code QFG, class 2.
As of Nov 19, 2026, K253743 is an FDA 510(k) held by Medtronic Minimed, Inc., substantially equivalent on Mar 12, 2026 after 108 days, with 1 recall tied to it in FDA's recall database.
- Review time: 108 days
- Recalls: 1
- MiniMed 780G insulin pumpK251032 · Medtronic Minimed, Inc. · Jul 1, 2025
| Started | Recall | Status |
|---|---|---|
| Aug 20, 2026 | Z-0088-2027Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The… | Open |
- Mint ACE PumpK263422 · Beta Bionics, Inc. · Oct 21, 2026
- Mint ACE PumpK260001 · Beta Bionics, Inc. · Sep 11, 2026
- twiist Automated Insulin Delivery (AID) SystemK261672 · Deka Research and Development · Jun 17, 2026
- iLet ACE PumpK253976 · Beta Bionics, Inc. · Apr 30, 2026
- MiniMed 780G Insulin Pump (MMT-1884); MiniMed 780G Insulin Pump Kit (MMT-1894)K253470 · Medtronic Minimed, Inc. · Jan 15, 2026
- Tandem Mobi insulin pump with interoperable technologyK253074 · Tandem Diabetes Care, Inc. · Nov 6, 2025
- iLet ACE PumpK252770 · Beta Bionics, Inc. · Sep 29, 2025
- t:slim X2 insulin pump with interoperable technologyK250792 · Tandem Diabetes Care, Inc. · Sep 24, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Nov 24, 2025
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 880.5730
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files