iLet ACE Pump
Alternate Controller Enabled Insulin Infusion Pump. 510(k) K252770, from Beta Bionics, Inc. (Concord, MA). Product code QFG, class 2.
As of Nov 19, 2026, K252770 is an FDA 510(k) held by Beta Bionics, Inc., substantially equivalent on Sep 29, 2025 after 27 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 27 days
- Recalls: 2
- Cited as a predicate by: 1 device
The iLet ACE Pump is an alternate controller enabled (ACE) pump intended to deliver insulin under the skin based on input from an integrated continuous glucose monitor (CGM) and an interoperable automated glycemic controller (iAGC), in people 6 years of age or older with diabetes mellitus. The iLet ACE Pump is intended for single-person use; it is not to be shared.
- iLet® ACE PumpK231485 · Beta Bionics, Inc. · Jun 21, 2023
- iLet ACE PumpK253976 · Beta Bionics, Inc. · Apr 30, 2026
| Mar 12, 2026 | Z-1996-2026insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous Glucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, leading to only basal/requested meal doses… | Open |
| Sep 23, 2025 | Z-2513-2026The reason for the recall is: it was found in version 3.3.6 of the iLet ACE Pump firmware and all previous releases that a screen unlock failure could occur when the user successfully slides the unlock slider on the… | Open |
- Mint ACE PumpK263422 · Beta Bionics, Inc. · Oct 21, 2026
- Mint ACE PumpK260001 · Beta Bionics, Inc. · Sep 11, 2026
- twiist Automated Insulin Delivery (AID) SystemK261672 · Deka Research and Development · Jun 17, 2026
- iLet ACE PumpK253976 · Beta Bionics, Inc. · Apr 30, 2026
- MiniMed Flex pumpK253743 · Medtronic Minimed, Inc. · Mar 12, 2026
- MiniMed 780G Insulin Pump (MMT-1884); MiniMed 780G Insulin Pump Kit (MMT-1894)K253470 · Medtronic Minimed, Inc. · Jan 15, 2026
- Tandem Mobi insulin pump with interoperable technologyK253074 · Tandem Diabetes Care, Inc. · Nov 6, 2025
- t:slim X2 insulin pump with interoperable technologyK250792 · Tandem Diabetes Care, Inc. · Sep 24, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Sep 2, 2025
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 880.5730
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files