NOxBOXi Nitric Oxide Delivery System
Apparatus, Nitric Oxide Delivery. 510(k) K233251, from Linde Gas & Equipment, Inc. (Livingston, TN). Product code MRN, class 2.
As of Nov 17, 2026, K233251 is an FDA 510(k) held by Linde Gas & Equipment, Inc., substantially equivalent on May 23, 2024 after 238 days, with 4 recalls tied to it in FDA's recall database.
- Substantially equivalent: May 23, 2024
- Review time: 238 days
- Recalls: 4
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFNOxBOXi Nitric Oxide Delivery System is intended for use by healthcare professionals for the delivery and monitoring of a constant (user set) concentration of nitric oxide (NO) and the monitoring of NO2 and O2 in the inspiratory ventilator lines of a patient undergoing nitric oxide therapy (iNO).
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- NOxBOXi Nitric Oxide Delivery SystemK231823 · Linde Gas & Equipment, Inc. · Aug 11, 2023
Recalls
Recalls FDA ties to K233251, newest first; each opens FDA's recall record| Sep 14, 2026 | Z-3306-2026Potential for irregular oxygen readings due to a manufacturing issue affecting certain oxygen sensors. | Open, Classified |
| Aug 19, 2026 | Z-3165-2026Potential for oxygen tubing leaks due to failure of the tubing or tubing connection, which may lead to oxygen desaturation. | Open, Classified |
| Sep 9, 2025 | Z-0029-2026The unexpected internal fault / system diagnostic error to result when a rapid succession of button selections is made to the user interface without waiting for the device to respond to user prompts. | Open, Classified |
| Sep 9, 2025 | Z-0028-2026Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have… | Open, Classified |
Other MRN Clearances
The newest 510(k)s for the same product code; their median review was 402 days; see every MRN device- ULSPIRA TS Nitric Oxide Therapy SystemK253095 · Airgas Therapeutics, LLC · Oct 30, 2026
- ULSPIRA TS Nitric Oxide Therapy SystemK242374 · Airgas Therapeutics, LLC · Jan 17, 2025
- EVOLVE Nitric Oxide Delivery SystemK242578 · Mallinckrodt Manufacturing, LLC · Jan 3, 2025
- EVOLVE Nitric Oxide Delivery SystemK240410 · Malinckrodt Manufacturing, LLC · May 28, 2024
- Evolve Nitric Oxide Delivery SystemK222930 · Mallinckrodt Manufacturing, LLC · Dec 6, 2023
- NOxBOXi Nitric Oxide Delivery SystemK231823 · Linde Gas & Equipment, Inc. · Aug 11, 2023
- Ulspira TS Nitric Oxide Therapy SystemK212409 · Airgas Therapeutics · Jun 30, 2023
- Inomax DSIRK211153 · Mallinckrodt Manufacturing, LLC · Apr 20, 2023
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Sep 28, 2023
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.5165
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files