NOxBOXi Nitric Oxide Delivery System

Apparatus, Nitric Oxide Delivery. 510(k) K231823, from Linde Gas & Equipment, Inc. (Livingston, TN). Product code MRN, class 2.

As of Nov 17, 2026, K231823 is an FDA 510(k) held by Linde Gas & Equipment, Inc., substantially equivalent on Aug 11, 2023 after 51 days, with 4 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Aug 11, 2023
  • Review time: 51 days
  • Recalls: 4
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

NOxBOXi Nitric Oxide Delivery System is intended for use by healthcare professionals for the delivery and monitoring of a constant (user set) concentration of nitric oxide (NO) and the monitoring of NO2 and O2 in the inspiratory ventilator lines of a patient undergoing nitric oxide therapy (iNO).

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K231823, newest first; each opens FDA's recall record
Sep 14, 2026Z-3306-2026Potential for irregular oxygen readings due to a manufacturing issue affecting certain oxygen sensors.Open, Classified
Aug 19, 2026Z-3165-2026Potential for oxygen tubing leaks due to failure of the tubing or tubing connection, which may lead to oxygen desaturation.Open, Classified
Sep 9, 2025Z-0029-2026The unexpected internal fault / system diagnostic error to result when a rapid succession of button selections is made to the user interface without waiting for the device to respond to user prompts.Open, Classified
Sep 9, 2025Z-0028-2026Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have…Open, Classified

Other MRN Clearances

The newest 510(k)s for the same product code; their median review was 402 days; see every MRN device

Record

Type
Special
Received
Jun 21, 2023
Review panel
Anesthesiology
Regulation
21 CFR 868.5165

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.