Ultrasound System 2300

System, Imaging, Pulsed Doppler, Ultrasonic. 510(k) K223830, from Bk Medical Aps (Herlev, DK). Product code IYN, class 2.

As of Nov 19, 2026, K223830 is an FDA 510(k) held by Bk Medical Aps, substantially equivalent on Apr 11, 2023 after 111 days, with 6 recalls tied to it in FDA's recall database.

  • Review time: 111 days
  • Recalls: 6
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Intended Use: The system is a diagnostic ultrasound imaging system used by qualified and trained healthcare professionals for ultrasound imaging, human body fluid flow analysis and puncture and biopsy guidance.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K223830, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Oct 9, 2025Z-0437-2026, Z-0439-2026, Z-0438-2026, Z-0440-2026GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general…Open
Sep 4, 2024Z-0136-2025The internal power connections at the power inlet of the ultrasound devices can come loose resulting in the potential for excessive heat which could lead to a thermal event or the device becoming inoperable.Open
Nov 30, 2023Z-0813-2024An issue on the Dual Live Compare feature on the bkActiv system could result in a measurement error.Open

Other IYN clearances

The newest 510(k)s for the same product code; their median review was 134 days; see every IYN device

Elsewhere

Record

Type
Traditional
Received
Dec 21, 2022
Review panel
Radiology
Regulation
21 CFR 892.1550

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files