Arca
Skip to content
Tools
Tools

5000 Compact Series Ultrasound Systems (Ultrasound System 5500 G, Ultrasound System 5500 P, Ultrasound System 5500 W, Ultrasound System 5500 CV, Ultrasound System 5300 G, Ultrasound System 5300 P, Ultrasound System 5300 W)

System, Imaging, Pulsed Doppler, Ultrasonic. 510(k) K222648, from Philips Ultrasound, LLC (Bothell, WA). Product code IYN, class 2.

As of Nov 20, 2026, K222648 is an FDA 510(k) held by Philips Ultrasound, LLC, substantially equivalent on Sep 27, 2022 after 26 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 26 days
  • Recalls: 2
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The 5000 Compact Series Ultrasound Systems (Ultrasound System 5500G, Ultrasound System 5500P, Ultrasound System 5500W, Ultrasound System 5500CV, Ultrasound System 5300G, Ultrasound System 5300P, Ultrasound System 5300W) are intended for diagnostic Ultrasound imaging in B (or 2D), 3D/4D, Color Doppler, Continuous Wave Doppler, Pulse Wave Doppler, Tissue Doppler, M-mode (including anatomical…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K222648, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 23, 2024Z-1757-2024Ultrasound system with: 1) transesophageal echocardiography transducer (TEE) connected/not selected 2) 1, 2 non-TEE connected 3) Imaging with non-TEE 4) Selected/not selected imaging transducer removed 5) Transducer…Open
Jun 30, 2023Z-2517-2023Transducer, an ultrasound system accessory, consists of two parts that are bonded together that may come apart due to chassis bonding issue, which may pose risk to users and patients of tissue damage, and electric shock…Completed

Other IYN clearances

The newest 510(k)s for the same product code; their median review was 134 days; see every IYN device

Record

Type
Traditional, third-party review
Received
Sep 1, 2022
Review panel
Radiology
Regulation
21 CFR 892.1550

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files