Blueprint Mixed Reality system
Orthopedic Stereotaxic Instrument. 510(k) K222510, from Tornier S.A.S. (Montbonnot-Saint-Martin, FR). Product code OLO, class 2.
As of Nov 20, 2026, K222510 is an FDA 510(k) held by Tornier S.A.S., substantially equivalent on Jan 20, 2023 after 154 days, with 10 recalls tied to it in FDA's recall database.
- Review time: 154 days
- Recalls: 10
- Cited as a predicate by: 1 device
Blueprint® Mixed Reality system is indicated for use during Total Shoulder Arthroplasty using Stryker’s FDA cleared implants that are also implants cleared for pre-operative planning with the Blueprint® Software. Blueprint® Mixed Reality system is intended to allow surgeons to visualize the Blueprint® 3D preoperative planning intra-operatively.
- ExactechGPS Total Shoulder ApplicationK173372 · Blue Ortho · Nov 24, 2017
- Incompass Total Ankle SystemK250037 · Wright Medical Technology, Inc. (Stryker) · Jun 20, 2025
| Mar 5, 2025 | Z-1540-2025, Z-1539-2025, Z-1534-2025, Z-1541-2025, Z-1538-2025, Z-1535-2025, Z-1536-2025, Z-1537-2025, Z-1542-2025Stryker identified an elevated complaint rate related to difficulties using the Blueprint Mixed Reality Guidance Instrumentation system is multifactorial, but in each instance prohibits the system from meeting its… | Open |
| Dec 18, 2024 | Z-0846-2025The Pin Guide is being recalled because the metal tube may disassemble from the plastic handle. This can result in the Blueprint Mixed Reality headset being unable to detect the Pin Guide and potentially prolonging… | Open |
- GPS AnkleK262567 · Blue Ortho · Oct 26, 2026
- 7D Surgical System (10-0001)K260403 · 7D Surgical ULC · Oct 19, 2026
- TMINI Miniature Robotic SystemK263234 · THINK Surgical, Inc. · Oct 16, 2026
- CarboClear® Cervical Navigated InstrumentsK263154 · CarboFix Orhtopedics , Ltd. · Oct 9, 2026
- Spine Planning (3.0); Spine & Trauma Navigation; Alignment System SpineK262257 · Brainlab SE · Sep 30, 2026
- Remex Spine Surgery Navigation - Universal InstrumentsK262506 · Remex Medical Corp. · Aug 20, 2026
- Navbit MARQ (PRD-001-SUP (Supine Registration Kit), PRD-001-LAT (Lateral Registration Kit))K261410 · Suite 201, National Innovation Centre · Jul 30, 2026
- SPINEART Navigation Instrument SystemK261205 · Spineart SA · Jul 29, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Aug 19, 2022
- Review panel
- Neurology
- Regulation
- 21 CFR 882.4560
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files