Arca
Skip to content
Tools
Tools

ExactechGPS Total Shoulder Application

Orthopedic Stereotaxic Instrument. 510(k) K173372, from Blue Ortho (Gieres, FR). Product code OLO, class 2.

As of Nov 20, 2026, K173372 is an FDA 510(k) held by Blue Ortho, substantially equivalent on Nov 24, 2017 after 28 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 28 days
  • Recalls: 1
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The ExactechGPS is intended for use during preoperative planning and during stereotaxic surgery to aid the surgeon in locating anatomical structures and aligning the endoprosthesis with the anatomical structures provided that the required anatomical landmarks can be identified on the patient's preoperative CT scan.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K173372, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Jun 5, 2023Z-2050-2023The ball and spring housed within the locking mechanism of the handle can dislodge during disassembly or during impaction.Open

Other OLO clearances

The newest 510(k)s for the same product code; their median review was 91 days; see every OLO device

Record

Type
Special
Received
Oct 27, 2017
Review panel
Neurology
Regulation
21 CFR 882.4560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files