ImaCor Zura Handheld ZHH-010

System, Imaging, Pulsed Doppler, Ultrasonic. 510(k) K220490, from Imacor, Inc. (Jericho, NY). Product code IYN, class 2.

As of Nov 19, 2026, K220490 is an FDA 510(k) held by Imacor, Inc., substantially equivalent on Apr 8, 2022 after 45 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 45 days
  • Recalls: 2

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The ImaCor Zura Handheld ZHH-010 is a software-based handheld ultrasound imaging system and accessories, intended for diagnostic imaging. It is indicated for diagnostic ultrasound imaging using the ImaCor ClariTEE probe. (Cardiac, TEE) The system is intended for use in critical care environments where healthcare is provided by trained healthcare professionals.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K220490, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 29, 2025Z-0038-2026Production assembly error resulting in the potential for reversed articulation, where moving the lever forward caused the probe tip to anteflex rather than retroflex, and vice versa.Open
Sep 16, 2024Z-0231-2025Epoxy used to seal the edge of the transducer into its paddle exceeded specification. This variability could attenuate the ultrasound signal with a potential impact on the quality of images obtained from the probes.Open

Other IYN clearances

The newest 510(k)s for the same product code; their median review was 134 days; see every IYN device

Record

Type
Traditional, third-party review
Received
Feb 22, 2022
Review panel
Radiology
Regulation
21 CFR 892.1550

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files