Arca
Skip to content
Tools
Tools

Puritan Bennett Cuff Pressure Manager

Cuff, Tracheal Tube, Inflatable. 510(k) K202874, from Covidien, LLC (Shanghai, CN). Product code BSK, class 2.

As of Nov 19, 2026, K202874 is an FDA 510(k) held by Covidien, LLC, substantially equivalent on Jan 9, 2021 after 103 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 103 days
  • Recalls: 1

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The CPM is intended to continuously measure and automatically maintain the cuff pressure of an endotracheal tube or tracheostomy tube that has been set by the user/clinician during mechanical ventilation.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • IntelliCuffK150893 · Hamilton Medical AG · Jan 29, 2016

Recalls

Recalls FDA ties to K202874, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Aug 22, 2023Z-2651-2023Inadequate internal fixation of the power inlet port to the cuff pressure manager device causes the external power inlet to became dislodged within the device housing after several uses, which prevents the device from…Open

Other BSK clearances

The newest 510(k)s for the same product code; their median review was 220 days; see every BSK device
  • TrachFlushK261499 · Aw Technologies Aps · Aug 13, 2026
  • CuffGuardK243562 · Idmed · Aug 12, 2025
  • CuffTrekK242642 · Shanghai Longmann Tech Co., Ltd. · Mar 7, 2025
  • TrachCuff Cuff ControllerK220854 · Aw Technologies Aps · Apr 14, 2023
  • AG100sK221477 · Hospitech Respiration , Ltd. · Dec 29, 2022

Record

Type
Traditional
Received
Sep 28, 2020
Review panel
Anesthesiology
Regulation
21 CFR 868.5750

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files