Cuff, Tracheal Tube, Inflatable. 510(k) K150893, from Hamilton Medical AG (Bonaduz, CH). Product code BSK, class 2.

As of Nov 19, 2026, K150893 is an FDA 510(k) held by Hamilton Medical AG, substantially equivalent on Jan 29, 2016 after 302 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 302 days
  • Recalls: 2
  • Cited as a predicate by: 4 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

K150893 510(k) Number: ____________________________ Device Name: IntelliCuff Indication for Use: The IntelliCuff device is intended to continuously measure and automatically maintain the user- set cuff pressure of an endotracheal tube (ETT) or tracheostomy tube (TT) during mechanical ventilation.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • PytonK092733 · Arm Medical Devices, Inc. · Feb 26, 2010

Cited as a predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K150893, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 22, 2026Z-2625-2026Due to reported complaints, the cuff device may alarm with a "Cuff System Leakage" error. As a result, the motor may pump continuously and fail to maintain the cuff pressure set by the user.Open
Jun 20, 2017Z-0335-2018Issue related to the performance of the motor in recalled product. During use, motor may cease to function.Open

Other BSK clearances

The newest 510(k)s for the same product code; their median review was 220 days; see every BSK device

Record

Type
Traditional
Received
Apr 2, 2015
Review panel
Anesthesiology
Regulation
21 CFR 868.5750

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files