DigitalDiagnost C90
Solid State X-Ray Imager (Flat Panel/Digital Imager). 510(k) K202564, from Philips Medical Systems Dmc GmbH (Hamburg, DE). Product code MQB, class 2.
As of Nov 20, 2026, K202564 is an FDA 510(k) held by Philips Medical Systems Dmc GmbH, substantially equivalent on Sep 30, 2020 after 26 days, with 8 recalls tied to it in FDA's recall database.
- Review time: 26 days
- Recalls: 8
- Cited as a predicate by: 2 devices
- DigitalDiagnost C90K182973 · Philips Medical Systems Dmc GmbH · Jan 11, 2019
- pixium 4343RCEK170113 · Philips Medical Systems Dmc GmbH · Feb 9, 2017
- Philips Prograde; Philips Eleva Workspot for Digital DiagnostK141736 · Philips Medical Systems Dmc GmbH · Jul 25, 2014
- Philips Eleva Workspot with SkyFlowK153318 · Philips Medical Systems Dmc GmbH · Dec 22, 2015
- Radiography 7300 CK233662 · Philips Medical Systems Dmc GmbH · Jun 13, 2024
- DigitalDiagnostK210692 · Philips Medical Systems Dmc GmbH · Apr 2, 2021
| May 1, 2026 | Z-2354-2026System was delivered without the required Seismic kit for Wall Stand VS2, which is designed to prevent damage and ensure stability of the Wall Stand VS2 during seismic events. | Open |
| Oct 1, 2025 | Z-0465-2026Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/physical exertion required for manual… | Open |
| Sep 21, 2023 | Z-0275-2024, Z-0276-2024Potential for units suspended on the ceiling to fail and fall. | Open |
| May 19, 2023 | Z-2042-2023There is a software login in issue that may prevent the user from logging in. | Open |
| Sep 22, 2021 | Z-0239-2022The Aleva monitor requires certain labels to be applied, some of which are required by FDA Electronic Product Radiation Control performance standards. This includes the Radiation Warning Label [21CFR 1020.30 (j)] and… | Open |
| May 19, 2021 | Z-1328-2022, Z-1329-2022Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of image on the first examination due to an issue in the… | Open |
- 0909FCE, 0909FCE-HSK253800 · Rayence Co., Ltd. · Jul 31, 2026
- Vivix-S 4386W (FXRD-4386WA)K260820 · Vieworks Co., Ltd. · Jun 18, 2026
- Universal 1417PIK252996 · Konica Minolta Healthcare Americas, Inc. · Jun 12, 2026
- Lux HD 2530 detector (Lux HD 2530)K252911 · Iray Imaging Technology (Haining) Limited · Oct 9, 2025
- Yushan X-Ray Flat Panel DetectorK250211 · Innocare Optoelectronics Corp. · Jul 22, 2025
- Skr 3000K250665 · Konica Minolta, Inc. · Jun 17, 2025
- Wireless/ Wired X-Ray Flat Panel DetectorsK243734 · Allengers Medical Systems Limited · Apr 18, 2025
- Expd 114; Expd 114G; Expd 114P; Expd 114PGK242770 · DRTECH Corporation · Mar 20, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Special, third-party review
- Received
- Sep 4, 2020
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1680
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files