Solid State X-Ray Imager (Flat Panel/Digital Imager). 510(k) K141736, from Philips Medical Systems Dmc GmbH (Hamburg, DE). Product code MQB, class 2.

As of Nov 17, 2026, K141736 is an FDA 510(k) held by Philips Medical Systems Dmc GmbH, substantially equivalent on Jul 25, 2014 after 28 days, with 6 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jul 25, 2014
  • Review time: 28 days
  • Recalls: 6
  • Cited as a predicate by: 3 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

As a part of a radiographic system, the Philips Eleva Workspot with SkyPlate Detectors is intended to acquire, process, store, display, and export digital radiographic images. The Philips Eleva Workspot with SkyPlate Detectors is suitable for all routine radiography exams, including specialist areas like intensive care, trauma, or pediatric work, excluding fluoroscopy, angiography and mammography.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K141736, newest first; each opens FDA's recall record
Sep 19, 2018Z-0239-2019Stitching run on SkyPlate detector interrupted intermittently; Stitching with SkyPlate aborts after first image; System software CD contains wrong baselineTerminated
Dec 11, 2017Z-0704-2018It the WiFi connection between the SkyPlate detector and HP transfer point is weak, an image may fail to transfer from the SkyPlate detector to the system. The image remains in the memory of the detector, but cannot be…Terminated
May 11, 2017Z-2154-2017During recent evaluations of the Philips DigitalDiagnost 3.1.x, we have identified a potential issue that may affect the performance of the equipment under certain conditions.Terminated
Nov 10, 2015Z-1695-2016The detector may signal that it is ready for acquisition when it actually is not, resulting in failure to properly acquire the X-ray image.Terminated
Mar 23, 2015Z-2383-2015The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal due to a defect in the Microsoft Windows 7…Terminated
Feb 9, 2015Z-1554-2015The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal. This is a failure to comply with 21CFR…Terminated

Other MQB Clearances

The newest 510(k)s for the same product code; their median review was 179 days; see every MQB device

Record

Type
Special
Received
Jun 27, 2014
Review panel
Radiology
Regulation
21 CFR 892.1680

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.