System, Planning, Radiation Therapy Treatment. 510(k) K200569, from RaySearch Laboratories AB (PUBL) (Stockholm, SE). Product code MUJ, class 2.

As of Nov 19, 2026, K200569 is an FDA 510(k) held by RaySearch Laboratories AB (PUBL), substantially equivalent on Sep 24, 2020 after 204 days, with 14 recalls tied to it in FDA's recall database.

  • Review time: 204 days
  • Recalls: 14
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • RayStationK190387 · RaySearch Laboratories AB (PUBL) · Jun 19, 2019

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K200569, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 8, 2025Z-2489-2025DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.Open
Jan 16, 2025Z-1187-2025Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT…Open
Mar 28, 2024Z-1745-2024, Z-1746-2024Potential for reported SSD to be too high.Open
Feb 23, 2024Z-1482-2024A use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.Open
Jun 1, 2022Z-1403-2022An issue with propagation of treatment course information from RayStation with RayTreat to RayCare has been identified. During some workflows, information may not be propagated to RayCare.Terminated
Jun 8, 2021Z-2138-2021Initial delivery positions will be set incorrectly when having setup beam(s) at the localization point.Terminated
Apr 21, 2021Z-1838-2021Evaluation of a clinical goal for the voxelwise worst dose distributions may be misleading after editing the clinical goal. It also describes an issue with contouring when the spacing between slices is d 1 mm, present…Terminated
Apr 7, 2021Z-1598-2021When using more than one plan in a treatment course, the Total dose display in Dose tracking may be incorrect. There may be a mismatch between the fractions that appear as selected in the Treatment course list in Dose…Terminated
Mar 25, 2021Z-1427-2021Flags edited in RayCare after a patient has been checked in may not always be updated in RayTreat.Open
Feb 22, 2021Z-1597-2021For some LINAC types, merging clinical beams with beams of approximate dose may lead to the approximate dose erroneously labeled as clinical dose. Merge beams can be used manually, in scripting, or as part of the…Terminated
Jan 22, 2021Z-2039-2023A software bug affecting results when using deep learning (DL) segmentation followed by geometrical functions: Create wall function, Expand Contract function or ROI algebra function, the results are incorrect. If dose…Terminated
Nov 11, 2020Z-0589-2021Two issues were found with the treatment plan report: 1) There may be incorrect information for the percent of a region of interest with a clinical goal that is outside the dose grid; and 2) For RayStation and RayPlan…Terminated
Oct 8, 2020Z-0478-2021If a region of interest (ROI) or point of interest (POI) that is referenced from an imported plan is missing in the imported RT Structure Set, the reference may become linked to the wrong ROI or POI.Terminated

Other MUJ clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every MUJ device

Elsewhere

Record

Type
Traditional
Received
Mar 4, 2020
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files