System, Planning, Radiation Therapy Treatment. 510(k) K190387, from RaySearch Laboratories AB (PUBL) (Stockholm, SE). Product code MUJ, class 2.

As of Nov 19, 2026, K190387 is an FDA 510(k) held by RaySearch Laboratories AB (PUBL), substantially equivalent on Jun 19, 2019 after 120 days, with 25 recalls tied to it in FDA's recall database.

  • Review time: 120 days
  • Recalls: 25+
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • RayStation 7K180379 · RaySearch Laboratories AB (PUBL) · Apr 25, 2018

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K190387, newest first; the 25 newest; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 8, 2025Z-2488-2025DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.Open
Jan 16, 2025Z-1186-2025Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT…Open
Mar 28, 2024Z-1744-2024, Z-1743-2024Potential for reported SSD to be too high.Open
Feb 23, 2024Z-1482-2024A use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.Open
Feb 17, 2023Z-1281-2023It is possible to set a non-zero collimator angle in the Virtual Simulation module. The angle will be correct in the user interface but if the plan is DICOM exported by using the VSIM Export button, the angle will not…Open
May 26, 2020Z-2426-2020It is not obvious that the manual import of a treatment record into RayCare PACS does not automatically import the same data into RayTreat. In addition, there is an issue found with the display of planned and recorded…Open
May 18, 2020Z-2346-2020, Z-2349-2020, Z-2353-2020, Z-2348-2020, Z-2344-2020, Z-2342-2020, Z-2340-2020, Z-2352-2020, Z-2339-2020, Z-2347-2020, Z-2350-2020, Z-2345-2020, Z-2343-2020, Z-2351-2020, Z-2341-2020Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.Terminated
Mar 4, 2020Z-1598-2020Due to physical limitations of the treatment machine, the maximum field opening for Siemens Virtual Wedge fields recommended by the LINAC vendor will be more restricted in the wedged direction than the maximum field…Open
Feb 20, 2020Z-1465-2020, Z-1467-2020There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the…Terminated

Other MUJ clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every MUJ device

Elsewhere

Record

Type
Traditional
Received
Feb 19, 2019
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files