RayCare 3.1

System, Planning, Radiation Therapy Treatment. 510(k) K200487, from RaySearch Laboratories AB (PUBL) (Stockholm, SE). Product code MUJ, class 2.

As of Nov 20, 2026, K200487 is an FDA 510(k) held by RaySearch Laboratories AB (PUBL), substantially equivalent on Jun 10, 2020 after 104 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 104 days
  • Recalls: 5
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

RayCare is an oncology information system used to support workflows and scheduling, as well as clinical information, planning and treatment management for oncology care. RayCare may be used in the transfer, storage, conversion and display of patient data in all areas of cancer care including radiation therapy, medical oncology and surgical oncology.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • RayCare 2.3K191384 · RaySearch Laboratories AB (PUBL) · Jul 8, 2019

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K200487, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Sep 12, 2023Z-0206-2024An issue was found in RayCare 5B, 6A, including service packs, where values will not be visible for documents in a read-only state when opened outside the Documents workspace when using document fields configured with…Open
Jul 17, 2023Z-2464-2023An issue was found in Software RayCare 5A, 5B, 6A, including service packs, where an allergy warning against medication substance (drug ingredient) will not be displayed as expected under certain circumstancesOpen
Jun 8, 2021Z-2138-2021Initial delivery positions will be set incorrectly when having setup beam(s) at the localization point.Terminated
Mar 3, 2021Z-1281-2021Patient related messages created in RayCare 3B, RayCare 4A will be lost when upgrading from RayCare 2 to RayCare 3B or laterOpen
Jul 29, 2020Z-2899-2020When performing offline image review in RayCare 2C, RayCare 3A and RayCare 3B an offline image review task will guide you to the image viewer with the treatment image related to the fraction and the planning image…Open

Other MUJ clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every MUJ device

Record

Type
Traditional
Received
Feb 27, 2020
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files