RAPIDPoint 500e Blood Gas System
Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph. 510(k) K192240, from Siemens Healthcare Diagnostics, Inc. (Norowood, MA). Product code CHL, class 2.
As of Nov 17, 2026, K192240 is an FDA 510(k) held by Siemens Healthcare Diagnostics, Inc., substantially equivalent on Mar 27, 2020 after 221 days, with 4 recalls tied to it in FDA's recall database.
- Substantially equivalent: Mar 27, 2020
- Review time: 221 days
- Recalls: 4
- Cited as a predicate by: 2 devices
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Measurement Of Ph In Pleural Fluid On Blood Gas SystemK122539 · Siemens Healthcare Diagnostics · Apr 19, 2013
Cited as a Predicate
2 later 510(k) summaries name this device as a predicate- i-STAT CG8+ cartridge with the i-STAT 1 SystemK230285 · Abbott Point of Care, Inc. · Oct 27, 2023
- i-STAT G3+ cartridge with the i-STAT 1 SystemK223857 · Abbott Point of Care, Inc. · Sep 15, 2023
Recalls
Recalls FDA ties to K192240, newest first; each opens FDA's recall record| May 17, 2024 | Z-2214-2024Siemens Healthineers has identified an issue with the RAPIDPoint500e system software version 5.3. This issue affects how the sample source is identified at the LIS when using the capillary mode. With this software… | Open, Classified |
| Sep 26, 2023 | Z-0422-2024There are potential drug interferences from Perhexiline Maleate or Atomoxetine Hydrochloride that may cause falsely elevated sodium results reported on the RAPIDPoint 500 and RAPIDPoint 500e Blood Gas Systems. The… | Open, Classified |
| Jun 1, 2022 | Z-1372-2022Discrepant (low bias) pO2 results, could be an incorrect diagnosis of hypoxemia and initiation of unnecessary or inappropriate respiratory support. | Open, Classified |
| May 27, 2020 | Z-2432-2020The scanner is not confirming the integrity of read barcode data using the check-digit when reading Code 39 barcodes. | Terminated |
Other CHL Clearances
The newest 510(k)s for the same product code; their median review was 267 days; see every CHL device- ABL90 FLEX PLUS System, safeCLINITUBESK252207 · Radiometer Medicals Aps · Apr 10, 2026
- i-STAT CG4+ cartridge with the i-STAT 1 SystemK244014 · Abbott Point of Care, Inc. · May 2, 2025
- ABL90 FLEX PLUS SystemK240998 · Radiometer Medicals Aps · Dec 13, 2024
- i-STAT CG8+ cartridge with the i-STAT 1 SystemK230285 · Abbott Point of Care, Inc. · Oct 27, 2023
- Stat Profile Prime Plus Analyzer SystemK221900 · Nova Biomedical Corporation · Sep 29, 2023
- i-STAT G3+ cartridge with the i-STAT 1 SystemK223857 · Abbott Point of Care, Inc. · Sep 15, 2023
- GEM Premier 7000 with IQM3K223608 · Instrumentation Laboratory Company · Aug 10, 2023
- EasyStat 300K211559 · Medica Corporation · Jan 27, 2023
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Aug 19, 2019
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1120
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files