Arca
Skip to content

RAPIDPoint 500e Blood Gas System

Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph. 510(k) K192240, from Siemens Healthcare Diagnostics, Inc. (Norowood, MA). Product code CHL, class 2.

As of Nov 17, 2026, K192240 is an FDA 510(k) held by Siemens Healthcare Diagnostics, Inc., substantially equivalent on Mar 27, 2020 after 221 days, with 4 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Mar 27, 2020
  • Review time: 221 days
  • Recalls: 4
  • Cited as a predicate by: 2 devices

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K192240, newest first; each opens FDA's recall record
May 17, 2024Z-2214-2024Siemens Healthineers has identified an issue with the RAPIDPoint500e system software version 5.3. This issue affects how the sample source is identified at the LIS when using the capillary mode. With this software…Open, Classified
Sep 26, 2023Z-0422-2024There are potential drug interferences from Perhexiline Maleate or Atomoxetine Hydrochloride that may cause falsely elevated sodium results reported on the RAPIDPoint 500 and RAPIDPoint 500e Blood Gas Systems. The…Open, Classified
Jun 1, 2022Z-1372-2022Discrepant (low bias) pO2 results, could be an incorrect diagnosis of hypoxemia and initiation of unnecessary or inappropriate respiratory support.Open, Classified
May 27, 2020Z-2432-2020The scanner is not confirming the integrity of read barcode data using the check-digit when reading Code 39 barcodes.Terminated

Other CHL Clearances

The newest 510(k)s for the same product code; their median review was 267 days; see every CHL device

Record

Type
Special
Received
Aug 19, 2019
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1120

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.