ETEST Eravacycline (ERV) (0.002 – 32 µg/mL)
Manual Antimicrobial Susceptibility Test Systems. 510(k) K192050, from Biomerieux S.A. (Marcy L'Etoile, FR). Product code JWY, class 2.
As of Nov 28, 2026, K192050 is an FDA 510(k) held by Biomerieux S.A., substantially equivalent on Sep 27, 2019 after 58 days, with 1 recall tied to it in FDA's recall database.
- Review time: 58 days
- Recalls: 1
- Cited as a predicate by: 1 device
Gram-negative and Gram-positive aerobic bacteria and fastidious bacteria. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC, in μg/mL) of different antimicrobial agents against microorganisms tested on agar media using overnight incubation.
- ETEST Telavancin (TLA) (0.002-32 ug/mL)K180936 · Biomerieux S.A. · Jul 3, 2018
- ETEST Fosfomycin (FO) (0.032-512 µg/mL)K210757 · Biomerieux S.A. · Nov 12, 2021
| Started | Recall | Status |
|---|---|---|
| Nov 9, 2022 | Z-0851-2023As the temperature and time out of range was exceeded, product performances cannot be guaranteed | Open |
- ETEST Gepotidacin (GEP) (0.016-256 µg/mL)K260447 · bioMérieux SA · Apr 21, 2026
- MTS Gepotidacin 0.016-256 µg/mLK260696 · Liofilchem s.r.l. · Apr 17, 2026
- MTS Aztreonam-Avibactam 0.016/4 - 256/4 µg/mLK252114 · Liofilchem s.r.l. · Oct 2, 2025
- The Sensititre 18-24 hour MIC or Breakpoint Susceptibility System with Colistin in the dilution range of 0.12-16 ug/mLK252062 · Thermo Fisher Scientific · Sep 25, 2025
- The Sensititre 18 - 24 hour MIC or Breakpoint Susceptibility System with Aztreonam/avibactam in the dilution range of…K250575 · Thermo Fisher Scientific · Sep 5, 2025
- The Sensititre 18-24 hour MIC or Breakpoint Susceptibility System with Amikacin in the dilution range of 0.25-256 ug/mLK252014 · Thermo Fisher Scientific · Aug 28, 2025
- MTS Sulbactam-Durlobactam 0.004/4 - 64/4 µg/mL (SUD)K251580 · Liofilchem s.r.l. · Aug 20, 2025
- The Sensititre 18-24 hour MIC or Breakpoint Susceptibility System with Ciprofloxacin in the dilution range of 0.002-64…K250990 · Thermo Fisher Scientific · Jun 30, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jul 31, 2019
- Review panel
- Microbiology
- Regulation
- 21 CFR 866.1640
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files