MTS Gepotidacin 0.016-256 µg/mL
Manual Antimicrobial Susceptibility Test Systems. 510(k) K260696, from Liofilchem s.r.l. (Roseto Degli Abruzzi, IT). Product code JWY, class 2.
As of Nov 28, 2026, K260696 is an FDA 510(k) held by Liofilchem s.r.l., substantially equivalent on Apr 17, 2026 after 45 days, with no recalls tied to it in FDA's recall database.
- Review time: 45 days
- Recalls: 0
The MTS (MIC Test Strip) Gepotidacin 0.016-256 μg/mL is a quantitative method intended for in vitro determination of antimicrobial susceptibility of bacteria. MTS consists of specialized paper impregnated with a predefined concentration gradient of an antimicrobial agent, which is used to determine the minimum inhibitory concentration (MIC) in μ g/mL of antimicrobial agents against bacteria as…
- Liofilchem MIC Test Strip (MTS)-Vancomycin 0.016 -256 ug/mLK153687 · Liofilchem s.r.l. · Mar 31, 2016
- ETEST Gepotidacin (GEP) (0.016-256 µg/mL)K260447 · bioMérieux SA · Apr 21, 2026
- MTS Aztreonam-Avibactam 0.016/4 - 256/4 µg/mLK252114 · Liofilchem s.r.l. · Oct 2, 2025
- The Sensititre 18-24 hour MIC or Breakpoint Susceptibility System with Colistin in the dilution range of 0.12-16 ug/mLK252062 · Thermo Fisher Scientific · Sep 25, 2025
- The Sensititre 18 - 24 hour MIC or Breakpoint Susceptibility System with Aztreonam/avibactam in the dilution range of…K250575 · Thermo Fisher Scientific · Sep 5, 2025
- The Sensititre 18-24 hour MIC or Breakpoint Susceptibility System with Amikacin in the dilution range of 0.25-256 ug/mLK252014 · Thermo Fisher Scientific · Aug 28, 2025
- MTS Sulbactam-Durlobactam 0.004/4 - 64/4 µg/mL (SUD)K251580 · Liofilchem s.r.l. · Aug 20, 2025
- The Sensititre 18-24 hour MIC or Breakpoint Susceptibility System with Ciprofloxacin in the dilution range of 0.002-64…K250990 · Thermo Fisher Scientific · Jun 30, 2025
- The Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC or Breakpoint Susceptibility System…K242843 · Thermo Fisher Scientific · Jun 20, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Mar 3, 2026
- Review panel
- Microbiology
- Regulation
- 21 CFR 866.1640
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files