Arca
Skip to content
Tools
Tools

EarlyVue VS30

Alarm, Blood-Pressure. 510(k) K190624, from Philips Medizin Systeme Boeblingen GmbH (Boeblingen, DE). Product code DSJ, class 2.

As of Nov 20, 2026, K190624 is an FDA 510(k) held by Philips Medizin Systeme Boeblingen GmbH, substantially equivalent on Oct 27, 2019 after 230 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 230 days
  • Recalls: 3
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The EarlyVue VS30 is intended to measure, display, alarm and, record physiological information of adult, pediatric and neonatal patients in hospitals and in out-of-hospital patient care settings in which care is administered by a healthcare professional (such as clinics, outpatient surgery facilities, long-term care facilities, and physician offices).

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K190624, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 31, 2023Z-2654-2023, Z-2655-2023A software issue prevents the monitor from alarming when the high/low acoustic respiration rate (RRa) limits are violated.Open
Mar 6, 2023Z-1645-2023EarlyVue VS30 Vital Signs Monitor and EarlyVue VS30 Vitals monitor Missing Calibration Alarm, a risk of an inaccurate CO2 measurement which may lead to a failure to recognize a change in patient condition.Open

Record

Type
Traditional
Received
Mar 11, 2019
Review panel
Cardiovascular
Regulation
21 CFR 870.1100

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files