Arca
Skip to content
Tools
Tools

SureSigns VS3 and 1 more

Alarm, Blood-Pressure. 510(k) K163649, from Philips Medical Systems (Andover, MA). Product code DSJ, class 2.

As of Nov 20, 2026, K163649 is an FDA 510(k) held by Philips Medical Systems, substantially equivalent on Mar 31, 2017 after 98 days, with 10 recalls tied to it in FDA's recall database.

  • Review time: 98 days
  • Recalls: 10
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Indications for Use for SureSigns VS3 Vital Signs MonitorThe SureSigns VS3 vital signs monitor is for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The SureSignsVS3 is for monitoring, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate
  • EarlyVue VS30K190624 · Philips Medizin Systeme Boeblingen GmbH · Oct 27, 2019

Recalls

Recalls FDA ties to K163649, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 19, 2019Z-1321-2019, Z-1320-2019, Z-1322-2019, Z-1317-2019, Z-1319-2019, Z-1323-2019, Z-1324-2019, Z-1318-2019System software inhibits the monitor and as a result does not measure, display and alarm for pulse rates above 240 beats per minute.Open
Jul 13, 2018Z-2894-2018, Z-2893-2018Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon…Open

Record

Type
Special
Received
Dec 23, 2016
Review panel
Cardiovascular
Regulation
21 CFR 870.1100

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files