Arca
Skip to content
Tools
Tools

DR 800 with Tomosynthesis

System, X-Ray, Tomographic. 510(k) K183275, from Agfa N.V. (Mortsel, BE). Product code IZF, class 2.

As of Nov 19, 2026, K183275 is an FDA 510(k) held by Agfa N.V., substantially equivalent on Feb 1, 2019 after 70 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 70 days
  • Recalls: 3
  • Cited as a predicate by: 4 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The DR 800 system is indicated for performing dynamic imaging examinations (fluoroscopy and/or rapid sequence) of the following anatomies/procedures: • Positioning fluoroscopy procedures • Gastro-intestinal examinations • Urogenital tract examinations • Angiography It is intended to replace fluoroscopic images obtained through image intensifier technology.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K183275, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 18, 2025Z-0923-2026It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.Open
Jul 29, 2024Z-2738-2024Potential for the front lever chain of the DR 800 table to fail.Open
Feb 10, 2020Z-1502-2020Tomosynthesis image acquisition sequence did not stop automatically after expected number of exposures.Terminated

Other IZF clearances

The newest 510(k)s for the same product code; their median review was 122 days; see every IZF device

Record

Type
Traditional
Received
Nov 23, 2018
Review panel
Radiology
Regulation
21 CFR 892.1740

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.