- Review time: 70 days
- Recalls: 3
- Cited as a predicate by: 4 devices
The DR 800 system is indicated for performing dynamic imaging examinations (fluoroscopy and/or rapid sequence) of the following anatomies/procedures: • Positioning fluoroscopy procedures • Gastro-intestinal examinations • Urogenital tract examinations • Angiography It is intended to replace fluoroscopic images obtained through image intensifier technology.
- Discovery XR656 With Volumerad (Digital Tomosynthesis)K132261 · Ge Medical Systems, LLC · Nov 18, 2013
- DR 800 with MUSICA DynamicK180589 · Agfa Healthcare N.V. · Apr 5, 2018
- DX-D Imaging Package (DR 10s), DX-D Imaging Package (DR 14s), DX-D Imaging Package (DR 14s)K161368 · Agfa Healthcare N.V. · Sep 22, 2016
- Nio 3MP-21K051901 · Barcoview · Aug 17, 2005
- Ortho Device, ADAPTIX 3D Orthopedic Imaging SystemK221949 · Pausch Medical GmbH · Jan 26, 2023
- ValoryK213469 · Agfa N.V. · Nov 18, 2021
- DR 800 with DSA, DR 800K212145 · Agfa N.V. · Aug 31, 2021
- DR 600 with TomosynthesisK193262 · Agfa N.V. · Mar 9, 2020
| Nov 18, 2025 | Z-0923-2026It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled. | Open |
| Jul 29, 2024 | Z-2738-2024Potential for the front lever chain of the DR 800 table to fail. | Open |
| Feb 10, 2020 | Z-1502-2020Tomosynthesis image acquisition sequence did not stop automatically after expected number of exposures. | Terminated |
- TAP2DK252649 · Nano-X Imaging , Ltd. · Feb 2, 2026
- Adaptix Ortho350K252133 · Adaptix Limited · Nov 6, 2025
- Nanox.ARC XK250850 · Nano-X Imaging , Ltd. · Apr 16, 2025
- Nanox.ARCK242395 · Nano-X Imaging , Ltd. · Dec 4, 2024
- Nanox.ARCK222934 · Nano-X Imaging , Ltd. · Apr 28, 2023
- Ortho Device, ADAPTIX 3D Orthopedic Imaging SystemK221949 · Pausch Medical GmbH · Jan 26, 2023
- CLISIS SYSTEMS, Discovery RF180K213081 · General Medical Merate S.P.A · Oct 26, 2021
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Nov 23, 2018
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1740
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files