System, X-Ray, Fluoroscopic, Image-Intensified. 510(k) K180589, from Agfa Healthcare N.V. (Mortsel, BE). Product code JAA, class 2.

As of Nov 19, 2026, K180589 is an FDA 510(k) held by Agfa Healthcare N.V., substantially equivalent on Apr 5, 2018 after 30 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 30 days
  • Recalls: 3
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The DR 800 system is indicated for performing dynamic imaging examinations (fluoroscopy and/or rapid sequence) of the following anatomies/procedures: • Positioning fluoroscopy procedures • Gastro-intestinal examinations • Urogenital tract examinations • Angiography It is intended to replace fluoroscopic images obtained through image intensifier technology.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K180589, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 29, 2024Z-2738-2024Potential for the front lever chain of the DR 800 table to fail.Open
Apr 6, 2020Z-2202-2020Under specific conditions (Fluoroscopic exam, ABS=OFF, manual change of parameters) wrong calculation of the dose/minute for fluoroscopy exams can be possibleTerminated
Jun 27, 2019Z-2056-2019The exposed area of fluo exams possibly does not match the active area of the detector within specified limits.Terminated

Other JAA clearances

The newest 510(k)s for the same product code; their median review was 110 days; see every JAA device

Elsewhere

Record

Type
Traditional
Received
Mar 6, 2018
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.