da Vinci SP Surgical System, EndoWrist SP Instruments, and Accessories

System, Surgical, Computer Controlled Instrument. 510(k) K173906, from Intuitive Surgical, Inc. (Sunnyvale, CA). Product code NAY, class 2.

As of Nov 20, 2026, K173906 is an FDA 510(k) held by Intuitive Surgical, Inc., substantially equivalent on May 31, 2018 after 160 days, with 9 recalls tied to it in FDA's recall database.

  • Review time: 160 days
  • Recalls: 9
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

GD9LQFL63Š Surgical System, Model SP1098: The ,QWXLWLYH6XUJLFDOŠ Endoscopic Instrument Control System (GD9LQFL63® Surgical System, Model SP1098) is intended to assist in the accurate control of ,QWXLWLYH6XUJLFDO(QGR:ULVW63Œ Instruments during urologic surgical procedures that are appropriate for a single port approach. The system is indicated for adult use.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K173906, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 22, 2026Z-2521-2026Operative components of surgical systems potentially distributed with E-brake retainer pin, so in event of a structural column failure, the E-brake pin would prevent E-brake engagement and may result in uncontrolled…Open
May 12, 2025Z-2722-2026Robotic-assisted surgical system arm drapes may detach from the instrument arm due to an issue involving the drape rings located at the top of the product, which may expose a section of instrument arm, which may result…Open
Mar 13, 2024Z-1518-2024Surgical system may have inadequately welded top and/or bottom belts in the link 2 of the entry guide manipulator assembly/instrument arm joints, that may cause the following, 1) Top Failure: internal tissue injuries,…Open
Sep 22, 2021Z-0419-2022Issue was identified during internal engineering evaluation. Use of the system with the affected software version may experience either an inability to deliver energy or an inadvertent delivery of energy. Use may…Open
Aug 19, 2019Z-2131-2020Some cameras were found to have residual soil adhered to the tip of the camera after reprocessing which may affect image quality resulting in needing to switch to a second camera or converting to another surgical…Open
Jul 26, 2019Z-0927-2020The firm become aware of a potentially defective capacitors on circuit boards installed on certain da Vinci Xi, X and SP system Vision Side Carts (VSC) and/or Patient Side Carts (PSC). As a result, you may experience…Terminated
Apr 1, 2019Z-2260-2020Intuitive has become aware that the da Vinci SP system may trigger a mechanical vibration of the instrument tips and endoscope due to a software anomaly. This issue has the potential to occur only under the following,…Terminated
Oct 25, 2018Z-1906-2020The instrument drape may tear while draping the Patient Side Cart presenting a breach in the sterile barrier.Terminated
Oct 23, 2018Z-0845-2020The firm became aware of a problem with the surgical system where the cannula mount button may be damaged when the user releases the button abruptly.Terminated

Other NAY clearances

The newest 510(k)s for the same product code; their median review was 89 days; see every NAY device

Elsewhere

Record

Type
Traditional
Received
Dec 22, 2017
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.1500

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files