Arca
Skip to content
Tools
Tools

EVIS EXERA III BRONCHOFIBERVIDEOSCOPE OLYMPUS BF-MP190F, Single Use Biopsy Forceps FB-433D

Bronchoscope (Flexible Or Rigid). 510(k) K172726, from Olympus Medical Systems Corp. (Hachioji-Shi, JP). Product code EOQ, class 2.

As of Nov 18, 2026, K172726 is an FDA 510(k) held by Olympus Medical Systems Corp., substantially equivalent on Apr 30, 2018 after 231 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Apr 30, 2018
  • Review time: 231 days
  • Recalls: 2
  • Cited as a predicate by: 3 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

EVIS EXERA III BRONCHOFIBERVIDEOSCOPE OLYMPUS BF-MP190F This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K172726, newest first; each opens FDA's recall record
Sep 25, 2023Z-0327-2024Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for…Open, Classified
Jul 27, 2023Z-2497-2023Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.Open, Classified

Other EOQ Clearances

The newest 510(k)s for the same product code; their median review was 229 days; see every EOQ device

Record

Type
Traditional
Received
Sep 11, 2017
Review panel
Ear, Nose, Throat
Regulation
21 CFR 874.4680

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.