Evis Exera Iii Video System
Bronchoscope (Flexible Or Rigid). 510(k) K121959, from Olympus Medical Systems Corp. (Center Valley, PA). Product code EOQ, class 2.
As of Nov 18, 2026, K121959 is an FDA 510(k) held by Olympus Medical Systems Corp., substantially equivalent on Jan 10, 2013 after 189 days, with 10 recalls tied to it in FDA's recall database.
- Substantially equivalent: Jan 10, 2013
- Review time: 189 days
- Recalls: 10
- Cited as a predicate by: 5 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFEndoscopes (OLYMPUS BF-Q190, OLYMPUS BF-H190, OLYMPUS BF-1TH190 EVIS EXERAIII BRONCHOVIDEOSCOPE) This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.
Cited as a Predicate
5 later 510(k) summaries name this device as a predicate- Evis Exera Iii Bronchovideoscope (Olympus BF-Q190); Evis Exera Iii Bronchovideoscope (Olympus BF-H190 ); Evis Exera Iii…K252341 · Olympus Medical Systems Corp. · Apr 23, 2026
- EVIS XI Video System Center Olympus CV-1500, Bronchovideoscope Olympus BF-H1100, Bronchovideoscope Olympus BF-1TH1100K222861 · Olympus Medical Systems Corporation · Jun 15, 2023
- EVIS EXERA III BRONCHOFIBERVIDEOSCOPE OLYMPUS BF-MP190F, Single Use Biopsy Forceps FB-433DK172726 · Olympus Medical Systems Corp. · Apr 30, 2018
- Ambu aScope 3 Regular 5.0/2.2, Ambu aScope 3 Slim 3.8/1.2, Ambu aScope 3 Large 5.8/2.8, Ambu aView MonitorK161656 · Ambu A/S · Sep 1, 2016
- PENTAX Medical EKP-i5010 Video Processor with EB FamilyK143727 · Pentax of America, Inc. · Oct 29, 2015
Recalls
Recalls FDA ties to K121959, newest first; each opens FDA's recall record| Sep 11, 2025 | Z-0050-2026Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment. | Open, Classified |
| Sep 11, 2025 | Z-0049-2026Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment. | Open, Classified |
| Sep 11, 2025 | Z-0063-2026Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment. | Open, Classified |
| Sep 11, 2025 | Z-0053-2026Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment. | Open, Classified |
| Sep 11, 2025 | Z-0060-2026Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment. | Open, Classified |
| Oct 12, 2023 | Z-0192-2024There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the… | Open, Classified |
| Sep 25, 2023 | Z-0327-2024Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for… | Open, Classified |
| Jul 27, 2023 | Z-2496-2023Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large. | Open, Classified |
| Jun 8, 2023 | Z-2015-2023There have been complaints of endobronchial combustion during therapeutic laser procedures with the bronchoscope. | Open, Classified |
| Jun 8, 2023 | Z-2017-2023There have been complaints of endobronchial combustion during therapeutic laser procedures with the bronchoscope. | Open, Classified |
Other EOQ Clearances
The newest 510(k)s for the same product code; their median review was 229 days; see every EOQ device- Single-use Flexible BronchoscopeK261637 · Hunan Vathin Medical Instrument Co., Ltd. · Sep 23, 2026
- Ambu® aScope 5 Broncho 2.7/1.2; Ambu® aScope 5 Broncho 4.2/2.2; Ambu® aScope 5 Broncho 4.2/2.2 Sampler SetK260309 · Ambu A/S · Sep 16, 2026
- 3.8mm Slim ProVu Video Bronchoscope (090-02-0038U); 5.0mm Regular ProVu Video Bronchoscope (090-02-0050U); 6.0mm Large…K254026 · Flexicare Medical Limited. · Sep 2, 2026
- MONARCH Platform (MON-000008)K260382 · Auris Health, Inc. · Jul 25, 2026
- LungFlow Airway CatheterK254037 · Free Flow Medical, Inc. · Jul 10, 2026
- FUJIFILM Endoscope Model EB-840S; FUJIFILM Endoscope Model EB-840TK253137 · Fujifilm Corporation · Jun 16, 2026
- SmartClip Delivery Catheter (ADR-1715)K260565 · Elucent Medical · May 14, 2026
- Evis Exera Iii Bronchovideoscope (Olympus BF-Q190); Evis Exera Iii Bronchovideoscope (Olympus BF-H190 ); Evis Exera Iii…K252341 · Olympus Medical Systems Corp. · Apr 23, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jul 5, 2012
- Review panel
- Ear, Nose, Throat
- Regulation
- 21 CFR 874.4680
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files