Arca
Skip to content

Evis Exera Iii Video System

Bronchoscope (Flexible Or Rigid). 510(k) K121959, from Olympus Medical Systems Corp. (Center Valley, PA). Product code EOQ, class 2.

As of Nov 18, 2026, K121959 is an FDA 510(k) held by Olympus Medical Systems Corp., substantially equivalent on Jan 10, 2013 after 189 days, with 10 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jan 10, 2013
  • Review time: 189 days
  • Recalls: 10
  • Cited as a predicate by: 5 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

Endoscopes (OLYMPUS BF-Q190, OLYMPUS BF-H190, OLYMPUS BF-1TH190 EVIS EXERAIII BRONCHOVIDEOSCOPE) This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.

Cited as a Predicate

5 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K121959, newest first; each opens FDA's recall record
Sep 11, 2025Z-0050-2026Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.Open, Classified
Sep 11, 2025Z-0049-2026Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.Open, Classified
Sep 11, 2025Z-0063-2026Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.Open, Classified
Sep 11, 2025Z-0053-2026Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.Open, Classified
Sep 11, 2025Z-0060-2026Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.Open, Classified
Oct 12, 2023Z-0192-2024There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the…Open, Classified
Sep 25, 2023Z-0327-2024Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for…Open, Classified
Jul 27, 2023Z-2496-2023Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.Open, Classified
Jun 8, 2023Z-2015-2023There have been complaints of endobronchial combustion during therapeutic laser procedures with the bronchoscope.Open, Classified
Jun 8, 2023Z-2017-2023There have been complaints of endobronchial combustion during therapeutic laser procedures with the bronchoscope.Open, Classified

Other EOQ Clearances

The newest 510(k)s for the same product code; their median review was 229 days; see every EOQ device

Record

Type
Traditional
Received
Jul 5, 2012
Review panel
Ear, Nose, Throat
Regulation
21 CFR 874.4680

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.