Bard Mission Disposable Core Biopsy Instrument
Instrument, Biopsy. 510(k) K171953, from Bard Peripheral Vascular, Inc. (Tempe, AZ). Product code KNW, class 2.
As of Nov 18, 2026, K171953 is an FDA 510(k) held by Bard Peripheral Vascular, Inc., substantially equivalent on Sep 14, 2017 after 77 days, with 3 recalls tied to it in FDA's recall database.
- Substantially equivalent: Sep 14, 2017
- Review time: 77 days
- Recalls: 3
- Cited as a predicate by: 4 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Bard® Mission® Disposable Core Biopsy Instrument is intended for use in obtaining biopsy samples from soft tissues such as lung, liver, spleen, kidney, prostate, lymph nodes, breast, thyroid, and various soft tissue tumors.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- M-Biopsy Semi-Automatic Biopsy Instrument M-Biopsy Coaxial Introducer Needle for SemK161409 · Mermaid Medical A/S · Aug 2, 2016
- Bard(R) Monopty(R) Disposable Core Biopsy Instrument, Bard(R) Max-Core(R) Disposable Core InstrumentK133948 · Bard Peripheral Vascular, Inc. · Feb 21, 2014
Cited as a Predicate
4 later 510(k) summaries name this device as a predicate- Marquee™ Disposable Core Biopsy Instrument and Instrument KitK250032 · Bard Peripheral Vascular, Inc. · Jul 10, 2025
- Automatic Core Biopsy Instrument/ Semi-Automatic Core Biopsy Instrument/ Disposable Coaxial Biopsy NeedleK241793 · Zhejiang Curaway Medical Technology Co., Ltd. · Jan 24, 2025
- M Biopsy /SureCore Automatic Disposable Biopsy Needle, M Biopsy /SureCore Semi-Automatic Disposable Biopsy Needle, M…K222865 · Canyon Medical, Inc. · Apr 3, 2023
- Temno Elite Biopsy SystemK201166 · Merit Medical Systems, Inc. · Aug 28, 2020
Recalls
Recalls FDA ties to K171953, newest first; each opens FDA's recall record| Nov 8, 2023 | Z-0643-2024The internal diameter of the coaxial cannula may be smaller or larger than the external diameter of the biopsy needle, and the length of the cannula may exceed the stated length on the label. As a result, the biopsy… | Open, Classified |
| Apr 11, 2023 | Z-1564-2023Coaxial cannula internal diameter and length of biopsy instruments exceed the external diameter and length of the biopsy needle, prohibiting the biopsy needle from properly fitting into the coaxial cannula and accessing… | Open, Classified |
| Feb 28, 2023 | Z-1260-2023The external diameter of the biopsy instrument is larger than the internal diameter of the coaxial needle, which could lead to: 1) Procedural delay, 2) Requirement of an additional device, 3), Needle tip breakage, if… | Open, Classified |
Other KNW Clearances
The newest 510(k)s for the same product code; their median review was 97 days; see every KNW device- OnControl Powered Bone Access Systems;OnControl Bone Marrow Aspiration System;OnControl Bone Marrow Biopsy…K262531 · Teleflex Medical, LLC · Oct 27, 2026
- Automatic Core Biopsy InstrumentK260517 · Zhejiang Curaway Medical Technology Co., Ltd. · Aug 28, 2026
- EnCor EnCompass™ Breast Biopsy and Tissue Removal SystemK262571 · Bard Peripheral Vascular, Inc. · Aug 14, 2026
- Brevera Biopsy System with CorLumina Imaging Technology (BREVDISP09)K262792 · Hologic, Inc. · Aug 12, 2026
- Marrow MarxmanK260572 · Medimetrix, LLC · Aug 4, 2026
- Promisemed VeriEcto Automatic Biopsy NeedlesK261196 · Promisemed Hangzhou Meditech Co., Ltd. · May 12, 2026
- Mammotome Prima™ MR Dual Vacuum-Assisted Breast Biopsy System Control Module; Mammotome Prima™ MR Dual Vacuum-Assisted…K260365 · Devicor Medical Products, Inc. · Mar 4, 2026
- ZeniCore™ Automatic Biopsy Instrument (863212100 / ZeniCore Automatic Biopsy Device 12ga x 10cm with Coaxial Introducer…K252795 · Argon Medical Devices, Inc. · Feb 23, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- Jun 29, 2017
- Review panel
- Gastroenterology, Urology
- Regulation
- 21 CFR 876.1075
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files