Bard Mission Disposable Core Biopsy Instrument

Instrument, Biopsy. 510(k) K171953, from Bard Peripheral Vascular, Inc. (Tempe, AZ). Product code KNW, class 2.

As of Nov 18, 2026, K171953 is an FDA 510(k) held by Bard Peripheral Vascular, Inc., substantially equivalent on Sep 14, 2017 after 77 days, with 3 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Sep 14, 2017
  • Review time: 77 days
  • Recalls: 3
  • Cited as a predicate by: 4 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Bard® Mission® Disposable Core Biopsy Instrument is intended for use in obtaining biopsy samples from soft tissues such as lung, liver, spleen, kidney, prostate, lymph nodes, breast, thyroid, and various soft tissue tumors.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K171953, newest first; each opens FDA's recall record
Nov 8, 2023Z-0643-2024The internal diameter of the coaxial cannula may be smaller or larger than the external diameter of the biopsy needle, and the length of the cannula may exceed the stated length on the label. As a result, the biopsy…Open, Classified
Apr 11, 2023Z-1564-2023Coaxial cannula internal diameter and length of biopsy instruments exceed the external diameter and length of the biopsy needle, prohibiting the biopsy needle from properly fitting into the coaxial cannula and accessing…Open, Classified
Feb 28, 2023Z-1260-2023The external diameter of the biopsy instrument is larger than the internal diameter of the coaxial needle, which could lead to: 1) Procedural delay, 2) Requirement of an additional device, 3), Needle tip breakage, if…Open, Classified

Other KNW Clearances

The newest 510(k)s for the same product code; their median review was 97 days; see every KNW device

Record

Type
Traditional, third-party review
Received
Jun 29, 2017
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.1075

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.