Bard(R) Monopty(R) Disposable Core Biopsy Instrument, Bard(R) Max-Core(R) Disposable Core Instrument

Instrument, Biopsy. 510(k) K133948, from Bard Peripheral Vascular, Inc. (Tempe, AZ). Product code KNW, class 2.

As of Nov 18, 2026, K133948 is an FDA 510(k) held by Bard Peripheral Vascular, Inc., substantially equivalent on Feb 21, 2014 after 60 days, with 7 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Feb 21, 2014
  • Review time: 60 days
  • Recalls: 7
  • Cited as a predicate by: 14 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

Similar penetration depth • Similar sample notch • Same number of samples • Same mechanics of action • Same mode of action • Same energy used / delivered • Same patient-contacting materials • Same fundamental scientific technology *Bard Peripheral Vascular, Inc.* *TRADE SECRET/CONFIDENTIAL INFORMATION Notify C. R.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

14 later 510(k) summaries name this device as a predicate; the 12 newest

Recalls

Recalls FDA ties to K133948, newest first; each opens FDA's recall record
Sep 30, 2024Z-0467-2025Disposable biopsy instruments were potentially manufactured missing the backend notch retention feature on the needle that results in inadequate securing of the needle inside the device, which could lead to needle…Open, Classified
Aug 10, 2023Z-0027-2024Diameter mismatch between coaxial and biopsy needle (cutting cannula) in disposable core biopsy instrument kits, where the coaxial cannula inner diameter prohibits the biopsy needle (cutting cannula) from properly…Open, Classified
Jan 14, 2021Z-1289-2021Due to the Penetration Depth Switch used to select the desired penetration depth of the biopsy needle for soft tissue biopsies, detaches from the device during use.Terminated
Oct 25, 2019Z-0724-2020During the firm's review of May 2019 complaint data, a significant increase in Disposable Core Biopsy Instrument complaints were observed for Failure to Fire, Firing Problem, and Failure to Obtain Samples.Terminated
Sep 22, 2017Z-0044-2018During the review of May 2017 complaint data, a significant increase in MAX-CORE Disposable Core Biopsy Instrument complaints was observed for Failure to Prime, Failure to Fire, Failure to Obtain Sample, and…Terminated
Apr 9, 2015Z-1586-2015The firm is recalling Bard MAX-CORE Disposable Core Biopsy Instrument due to the risk of having self-activation-related issues.Terminated
Mar 17, 2015Z-1447-2015Bard Peripheral Vascular is recalling the Bard Monopty Disposable Core Biopsy Instruments and Kits because they may be at risk of having activation-related issues.Terminated

Other KNW Clearances

The newest 510(k)s for the same product code; their median review was 97 days; see every KNW device

Record

Type
Special
Received
Dec 23, 2013
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.1075

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files