da Vinci X Surgical System
System, Surgical, Computer Controlled Instrument. 510(k) K171294, from Intuitive Surgical, Inc. (Sunnyvale, CA). Product code NAY, class 2.
As of Nov 18, 2026, K171294 is an FDA 510(k) held by Intuitive Surgical, Inc., substantially equivalent on May 26, 2017 after 24 days, with 10 recalls tied to it in FDA's recall database.
- Substantially equivalent: May 26, 2017
- Review time: 24 days
- Recalls: 10
- Cited as a predicate by: 17 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System, Model IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopics, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation…
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Da Vinci Surgical System, Endowrist Instruments And AccessoriesK131861 · Intuitive Surgical, Inc. · Mar 28, 2014
- Intuitive Surgical Da Vinci Si Surgical System: Model IS3000K081137 · Intuitive Surgical, Inc. · Feb 18, 2009
Cited as a Predicate
17 later 510(k) summaries name this device as a predicate; the 12 newest- 8mm Monopolar Curved Scissors (470179)K242610 · Iconocare Health · Mar 11, 2025
- da Vinci X Surgical System (IS4200); da Vinci Xi Surgical System (IS4000)K240852 · Intuitive Surgical, Inc. · Jun 11, 2025
- da Vinci Xi Surgical System (IS4000); da Vinci X Surgical System (IS4200)K240723 · Intuitive Surgical · May 24, 2024
- da Vinci Xi Surgical System (IS4000), da Vinci X Surgical System (IS4200)K231224 · Intuitive Surgical, Inc. · May 12, 2023
- da Vinci X Surgical System (IS4200), da Vinci Xi Surgical System (IS4000)K223080 · Intuitive Surgical, Inc. · Nov 22, 2022
- 8mm Monopolar Curved ScissorsK220023 · Intuitive Surgical, Inc. · Jan 31, 2022
- da Vinci Xi Surgical System (IS4000), da Vinci X Surgical System (IS4200)K211784 · Intuitive Surgical, Inc. · Aug 6, 2021
- da Vinci Xi Surgical System (IS4000), da Vinci X Surgical System (IS4200)K202834 · Intuitive Surgical, Inc. · Dec 10, 2020
- da Vinci Xi Surgical System, da Vinci X Surgical SystemK192803 · Intuitive Surgical · Apr 29, 2020
- da Vinci Xi Surgical System, da Vinci X Surgical SystemK191529 · Intuitive Surgical · Feb 6, 2020
- da Vinci X and Xi Surgical SystemK183086 · Intuitive Surgical, Inc. · Mar 31, 2020
- da Vinci Xi Surgical System, da Vinci X Surgical SystemK182140 · Intuitive Surgical, Inc. · Oct 24, 2018
Recalls
Recalls FDA ties to K171294, newest first; each opens FDA's recall record| Apr 2, 2026 | Z-2046-2026Due to manipulator containing screws with in the arm sub-assembly that my be susceptible to breaking. | Open, Classified |
| Nov 7, 2025 | Z-1232-2026Due to a software implementation error that allowed instruments arms failing an important diagnostic test to remain in clinical use. This test was supposed to be used to detect Universal Surgical Manipulator insertion… | Open, Classified |
| Feb 10, 2023 | Z-1293-2023Due to Universal Surgical Manipulator (USM) instrument carriage being loose as a result of the linear rail component not being manufactured within specifications. | Open, Classified |
| Jan 24, 2023 | Z-1262-2023Preventative maintenance data was used to identify instrument Arms, part of surgical systems, that allow surgeons to perform surgical tasks, that may be at risk of insertion ballscrew failure, allowing for uncontrolled… | Open, Classified |
| Sep 20, 2022 | Z-0133-2023Inadvertent energy delivery from surgical system instrument if 1) Force bipolar and bipolar instruments installed on system 2) Force bipolar connected to Force Triad generator, bipolar not connected to generator 3)… | Open, Classified |
| Jul 26, 2019 | Z-0926-2020The firm become aware of a potentially defective capacitors on circuit boards installed on certain da Vinci Xi, X and SP system Vision Side Carts (VSC) and/or Patient Side Carts (PSC). As a result, you may experience… | Terminated |
| Apr 11, 2019 | Z-1897-2020Failure of Universal Surgical Manipulators due to a high friction points, resulting in an increase of arm movement to include: feelings of resistance, hesitation, and jerking. | Terminated |
| Nov 21, 2018 | Z-1961-2020Due to a manufacturing variation in the arm that can result in a persistent recoverable fault error (error 23087) and loss of arm functionality prior or during surgery. | Terminated |
| Jul 16, 2018 | Z-3254-2018Certain Redundant Medical Grade Power Supply (RMGPS) units specific to Surgeon Console and Patient Side Carts were manufactured incorrectly and may fail prior or during use. | Terminated |
| Mar 7, 2018 | Z-2341-2018Specific da Vinci X Surgical Systems (IS4200) were shipped with defective Arm Controller Joint (ACJ) boards that were found to have a high rate of failure due to a defective component. | Terminated |
Other NAY Clearances
The newest 510(k)s for the same product code; their median review was 89 days; see every NAY device- da Vinci Xi Surgical System (IS4000);da Vinci X Surgical System (IS4200)K263592 · Intuitive Surgical, Inc. · Oct 30, 2026
- da Vinci Surgical System, Model IS5000K261823 · Intuitive Surgical, Inc. · Sep 8, 2026
- da Vinci Surgical System, Model IS5000K261824 · Intuitive Surgical, Inc. · Jul 31, 2026
- da Vinci Firefly Imaging SystemK260695 · Intuitive Surgical · May 22, 2026
- Senhance Ultrasonic SystemK254192 · Asensus Surgical · May 12, 2026
- da Vinci Force Feedback InstrumentsK253986 · Intuitive Surgical, Inc. · Mar 11, 2026
- da Vinci Surgical System (IS5000)K251739 · Intuitive Surgical, Inc. · Jan 16, 2026
- da Vinci SP Surgical System (SP1098); da Vinci E-200 GeneratorK253556 · Intuitive Surgical, Inc. · Jan 16, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- May 2, 2017
- Review panel
- Gastroenterology, Urology
- Regulation
- 21 CFR 876.1500
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files