CADD®- Solis Ambulatory Infusion Pump with Wireless Communication and 4 more

Pump, Infusion, Pca. 510(k) K170982, from Smiths Medical Ads, Inc. (Plymouth, MN). Product code MEA, class 2.

As of Nov 20, 2026, K170982 is an FDA 510(k) held by Smiths Medical Ads, Inc., substantially equivalent on Aug 24, 2017 after 143 days, with 6 recalls tied to it in FDA's recall database.

  • Review time: 143 days
  • Recalls: 6
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

CADD®- Solis Ambulatory Infusion Pump with Wireless Communication CADD®-Solis Infusion Pump, Model 2110 The CADD®-Solis Pump is indicated for intravenous, intra-arterial, subcutaneous, intraperitoneal, in close proximity to nerves, into an intraoperative site (soft tissue, body cavity/surgical wound site), epidural space or subarachnoid space.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K170982, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 10, 2025Z-1681-2025There is a potential for thermal damage in CADD-Solis and CADD- Solis VIP infusion pumps.Open
Apr 10, 2025Z-1667-2025Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.Open
Apr 10, 2025Z-1662-2025Pumps may experience Wireless Connection Modules intermittent connection alarms, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.Open
Jun 26, 2024Z-0466-2025The adapter input plug can become damaged or broken. If the input plug is damaged, the metal contacts to the body of the AC adapter may be exposed or one or more of the AC mains prongs may separate from the input plug.Open
Feb 27, 2024Z-2420-2024The CADD Solis VIP Ambulatory Infusion Pump is indicated for the following uses: " For intravenous, intraarterial, subcutaneous, intraperitoneal, perineural, surgical site, epidural space, or subarachnoid space infusion.Open
Dec 9, 2022Z-0963-2023There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" AlarmsOpen

Record

Type
Traditional
Received
Apr 3, 2017
Review panel
General Hospital
Regulation
21 CFR 880.5725

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files