Cadd Administration Set With Flow Stop, Medication Cassette Reservoir With Flow Stop and Extension Set

Set, Administration, Intravascular. 510(k) K040636, from Smiths Medical MD, Inc. (St. Paul, MN). Product code FPA, class 2.

As of Nov 19, 2026, K040636 is an FDA 510(k) held by Smiths Medical MD, Inc., substantially equivalent on May 19, 2004 after 70 days, with 6 recalls tied to it in FDA's recall database.

  • Review time: 70 days
  • Recalls: 6
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

"The CADD® Medication Cassette Reservoir with Flow Stop is designed for use with the CADD® pumps (except CADD-Micro® and CADD-TPN®) for delivery of medications and fluids." "The CADD® Administration Set with Flow Stop is designed for use with the CADD® pumps (except CADD-Micro® and CADD-TPN®) for delivery of medications and fluids."

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K040636, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 6, 2025Z-2522-2025Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.Open
Dec 9, 2022Z-0963-2023There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" AlarmsOpen
Mar 17, 2017Z-1767-2017Smiths Medical has become aware that the drawing on the Instructions for Use (IFU) on certain yellow striped CADD Administration Sets (part number 21-7339-24), incorrectly represent the actual filter size of the…Terminated
Apr 22, 2016Z-1831-2016Smiths Medical became aware that under delivery of medication may occur on CADD Administration Sets with Flow Stop.Terminated
Feb 2, 2015Z-1286-2015Smiths Medical has become aware of an issue with specific lots of 50mL CADD" Medication Cassette Reservoirs (Cassette). Some Cassettes may leak at the sealing area of the pump tube and medication bag. Smiths Medical has…Terminated
Dec 14, 2007Z-0879-2008Leakage associated with the CADD Medication Cassette Reservoirs for use with the CADD Ambulatory Infusion PumpsTerminated

Other FPA clearances

The newest 510(k)s for the same product code; their median review was 266 days; see every FPA device

Record

Type
Traditional
Received
Mar 10, 2004
Review panel
General Hospital
Regulation
21 CFR 880.5440

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files