Trinity™ Dual Mobility System
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented. 510(k) K170359, from Corin U.S.A. Limited (Tampa, FL). Product code LZO, class 2.
As of Nov 18, 2026, K170359 is an FDA 510(k) held by Corin U.S.A. Limited, substantially equivalent on Oct 31, 2017 after 267 days, with 1 recall tied to it in FDA's recall database.
- Review time: 267 days
- Recalls: 1
- Cited as a predicate by: 3 devices
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe Trinity™ Dual Mobility System is intended for use in the following indications: 1. Non-inflammatory degenerative joint disease, including osteoarthritis & avascular necrosis 2. Rheumatoid Arthritis 3. Correction of functional deformity 4. Revision of previously failed total hip arthroplasty, 5. Patients at increased risk of dislocation 6.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Modular Dual Mobility Liner, X3 Acetabular InsertK103233 · Howmedica Osteonics Corp. · Feb 3, 2011
- G7 Dual Mobility SystemK150522 · Biomet, Inc. · May 1, 2015
- Corin Trinity Acetbular System With Hxlpe LinersK110087 · Corin USA · May 3, 2011
- Corin Trinity Acetabular System Ecima LinersK111481 · Corin USA · Feb 6, 2012
- Trinity Acetabular SystemK131647 · Corin USA · Sep 27, 2013
Cited as a predicate
3 later 510(k) summaries name this device as a predicate- Momentum Acetabular Cup SystemK253948 · United Orthopedic Corporation · Sep 2, 2026
- Exactech Alteon Modular Dual Mobility (MDM) SystemK190890 · Exactech, Inc. · Oct 22, 2019
- BiMobile Dual Mobility System - E-Dur InsertsK190535 · Waldemar Link GmbH & Co. KG · Aug 6, 2019
Recalls
Recalls FDA ties to K170359, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Started | Recall | Status |
|---|---|---|
| Jul 28, 2021 | Z-2608-2021The packaging system on the devices have potential physical and water damage resulting in damage to the devices, loss of sterility, or contamination of the devices. | Terminated |
Other LZO clearances
The newest 510(k)s for the same product code; their median review was 102 days; see every LZO device- NEXUS® Hip StemK262473 · Microport Orthopedics, Inc. · Oct 1, 2026
- M-Vizion Femoral Revision System ExtensionK260173 · Medacta International S.A. · Sep 17, 2026
- Momentum Acetabular Cup SystemK253948 · United Orthopedic Corporation · Sep 2, 2026
- MobileLink Acetabular Cup System – 40mm ID Inserts (various)K261009 · Waldemar Link GmbH & Co. KG · Jul 6, 2026
- Z1 Hip SystemK260831 · Zimmer, Inc. · Jun 16, 2026
- Avenir® Müller Stem; Avenir Complete™ Hip SystemK260182 · Zimmer, Inc. · Apr 27, 2026
- implantcast Packaging System UpdateK260037 · Implantcast GmbH · Mar 30, 2026
- implaFit® short stemsK252401 · Implantcast GmbH · Mar 12, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Feb 6, 2017
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3353
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files