Arca
Skip to content
Tools
Tools

Signa Pet/Mr

Tomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance. 510(k) K163619, from Ge Medical Systems, LLC (Waukesha, WI). Product code OUO, class 2.

As of Nov 21, 2026, K163619 is an FDA 510(k) held by Ge Medical Systems, LLC, substantially equivalent on Feb 14, 2017 after 54 days, with 7 recalls tied to it in FDA's recall database.

  • Review time: 54 days
  • Recalls: 7
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The SIGNA PET/MR system combines magnetic resonance diagnostic devices (MRDD) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information, acquired simultaneously and isocentrically.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • Signa Pet/MrK142098 · Ge Medical Systems, LLC · Nov 24, 2014

Cited as a predicate

1 later 510(k) summary names this device as a predicate
  • Signa Pet/MrK213709 · Ge Medical Systems, LLC · Jan 20, 2022

Recalls

Recalls FDA ties to K163619, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jun 20, 2025Z-2309-2025GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T,…Open
Oct 28, 2024Z-0520-2025Gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning.Open
Jun 14, 2024Z-2466-2024GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE…Open
Dec 23, 2021Z-0595-2022The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.Open
Jun 4, 2021Z-2057-2021GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes…Open
Sep 8, 2020Z-0131-2021Potential electrical arc flash within the Power Gradient & RF (PGR) Cabinet.Terminated
May 11, 2017Z-3029-2017Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.Terminated

Other OUO clearances

The newest 510(k)s for the same product code; their median review was 147 days; see every OUO device
  • uPMR ZenithK261676 · Shanghai United Imaging Healthcare Co., Ltd. · Oct 28, 2026
  • Cubresa BrainPETK253963 · Cubresa, Inc. · Jul 17, 2026
  • BIOGRAPH OneK253023 · Siemens Healthineers AG · Jan 15, 2026
  • uPMR 790K234154 · Shanghai United Imaging Healthcare Co., Ltd. · May 24, 2024
  • uPMR 790K222540 · Shanghai United Imaging Healthcare Co., Ltd. · Nov 14, 2022
  • Signa Pet/MrK213709 · Ge Medical Systems, LLC · Jan 20, 2022

Record

Type
Traditional
Received
Dec 22, 2016
Review panel
Radiology
Regulation
21 CFR 892.1200

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files