Signa Pet/Mr
Tomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance. 510(k) K142098, from Ge Medical Systems, LLC (Waukesha, WI). Product code OUO, class 2.
As of Nov 21, 2026, K142098 is an FDA 510(k) held by Ge Medical Systems, LLC, substantially equivalent on Nov 24, 2014 after 115 days, with 6 recalls tied to it in FDA's recall database.
- Review time: 115 days
- Recalls: 6
- Cited as a predicate by: 2 devices
The SIGNA PET/MR system combines magnetic resonance diagnostic devices (MRDD) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information, acquired simultaneously and isocentrically.
- Biograph MmrK133226 · Siemens Medical Solutions USA, Inc. · Nov 12, 2013
- uPMR 790K183014 · Shanghai United Imaging Healthcare Co., Ltd. · Feb 27, 2019
- Signa Pet/MrK163619 · Ge Medical Systems, LLC · Feb 14, 2017
| Jun 20, 2025 | Z-2309-2025GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T,… | Open |
| Oct 28, 2024 | Z-0520-2025Gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. | Open |
| Jun 14, 2024 | Z-2466-2024GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE… | Open |
| Dec 23, 2021 | Z-0595-2022The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements. | Open |
| May 11, 2017 | Z-3029-2017Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard. | Terminated |
| Aug 31, 2015 | Z-0260-2016When performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR… | Terminated |
- uPMR ZenithK261676 · Shanghai United Imaging Healthcare Co., Ltd. · Oct 28, 2026
- Cubresa BrainPETK253963 · Cubresa, Inc. · Jul 17, 2026
- BIOGRAPH OneK253023 · Siemens Healthineers AG · Jan 15, 2026
- uPMR 790K234154 · Shanghai United Imaging Healthcare Co., Ltd. · May 24, 2024
- uPMR 790K222540 · Shanghai United Imaging Healthcare Co., Ltd. · Nov 14, 2022
- Signa Pet/MrK213709 · Ge Medical Systems, LLC · Jan 20, 2022
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Aug 1, 2014
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1200
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files