Display, Cathode-Ray Tube, Medical. 510(k) K162012, from GE Healthcare (Milwaukee, WI). Product code DXJ, class 2.

As of Nov 17, 2026, K162012 is an FDA 510(k) held by GE Healthcare, substantially equivalent on Sep 15, 2016 after 56 days, with 6 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Sep 15, 2016
  • Review time: 56 days
  • Recalls: 6
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The CARESCAPE Central Station is intended for use under the direct supervision of a licensed healthcare practitioner. The intended use is to provide clinicians with adult, pediatric and neonatal patient data within a hospital or clinical environment.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K162012, newest first; each opens FDA's recall record
Jul 30, 2026Z-3143-2026Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.Open, Classified
Sep 5, 2025Z-0214-2026Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only…Open, Classified
Sep 5, 2025Z-0213-2026Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only…Open, Classified
May 10, 2022Z-1265-2022The CARESAPE Central Station (CSCS) V2 can shut down due to a potential power supply component failure. This can lead to loss of patient monitoring at the central station.Open, Classified
Jan 28, 2022Z-0808-2022If the CARESCAPE Central Station v2.0 is used with an unapproved keyboard, the audio can be muted resulting in loss of audible alarms.Open, Classified
Jan 11, 2018Z-0487-2018CARESCAPE Central Station (CSCS) software version 2.0.2 units may experience unexpected NO COMM (No Communication) and network communication issues after boot-up or system restart.Terminated

Other DXJ Clearances

The newest 510(k)s for the same product code; their median review was 135 days; see every DXJ device

Record

Type
Special
Received
Jul 21, 2016
Review panel
Cardiovascular
Regulation
21 CFR 870.2450

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files