Arca
Skip to content
Tools
Tools

Uroview FD

System, X-Ray, Fluoroscopic, Image-Intensified. 510(k) K161019, from Pausch Medical GmbH (Erlangen, DE). Product code JAA, class 2.

As of Nov 20, 2026, K161019 is an FDA 510(k) held by Pausch Medical GmbH, substantially equivalent on Aug 17, 2016 after 127 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 127 days
  • Recalls: 1
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Uroview FD is a solid state detector fluoroscopic X-ray system, primarily for urological applications (functional x- ray-diagnostics, endourology and minimal invasive urology/surgery). The system, which includes a radiologic/ urologic treatment table, may be used for urological treatment, planning and diagnostic procedures including but not limited to: • Querying and retrieving patient…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • Uroskop OmniaK101491 · Siemens Medical Solutions USA, Inc. · Jun 30, 2010

Cited as a predicate

2 later 510(k) summaries name this device as a predicate
  • XR-MX/1000K200965 · Karl Storz Endoscopy America, Inc. · Dec 29, 2020
  • Uroview FD IIK180651 · Pausch Medical GmbH · Apr 12, 2018

Recalls

Recalls FDA ties to K161019, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Mar 14, 2017Z-2152-2017During patient interventions with the urology table the radiation is activated with a footswitch connected to the table. This footswitch is actively involved in the safety circuit of the urological table. Only if…Terminated

Other JAA clearances

The newest 510(k)s for the same product code; their median review was 110 days; see every JAA device

Record

Type
Traditional
Received
Apr 12, 2016
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files