ABL90 Flex Plus
Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph. 510(k) K160153, from Radiometer Medical Aps (Broenshoej, DK). Product code CHL, class 2.
As of Nov 17, 2026, K160153 is an FDA 510(k) held by Radiometer Medical Aps, substantially equivalent on Nov 4, 2016 after 287 days, with 5 recalls tied to it in FDA's recall database.
- Substantially equivalent: Nov 4, 2016
- Review time: 287 days
- Recalls: 5
- Cited as a predicate by: 3 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe ABL90 FLEX PLUS analyzer is an in vitro diagnostic, portable, automated analyser that quantitatively measures, pH, blood gases, electrolytes, glucose, lactate and oximetry in heparinized whole blood, and neonatal bilirubin in heparinized capillary whole blood. The ABL90 FLEX PLUS analyzer is intended for use by trained technologists, nurses, physicians and therapists.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- ABL90 FlexK132691 · Radiometer Medical Aps · Nov 13, 2014
- ABL90 Flex AnalyzerK131988 · Radiometer Medical Aps · Sep 17, 2013
Cited as a Predicate
3 later 510(k) summaries name this device as a predicate- ABL90 FLEX PLUS System, safeCLINITUBESK252207 · Radiometer Medicals Aps · Apr 10, 2026
- ABL90 FLEX PLUS SystemK240998 · Radiometer Medicals Aps · Dec 13, 2024
- ABL90 Flex, ABL90 Flex PlusK170882 · Radiometer Medical Aps · Apr 28, 2017
Recalls
Recalls FDA ties to K160153, newest first; each opens FDA's recall record| Mar 21, 2025 | Z-1659-2025ABL90 FLEX and ABL90 FLEX PLUS analyzers running with Radiometer Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze (the analyzer ceases to operate) during a sample measurement. The analyzer software… | Open, Classified |
| Apr 14, 2021 | Z-1858-2021Due to an issue with the internal clock of the analyzers with an e3800 PC unit installed. | Open, Classified |
| Apr 14, 2021 | Z-1857-2021Due to an issue with the internal clock of the analyzers with an e3800 PC unit installed. | Open, Classified |
| Feb 3, 2020 | Z-1761-2020The firm received reports of occurrences where the barcode reader misinterpreted the contents of a locally printed barcode label used for entering patient ID or accession number into the analyzer in connection with a… | Terminated |
| Nov 21, 2019 | Z-2138-2020The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has… | Terminated |
Other CHL Clearances
The newest 510(k)s for the same product code; their median review was 267 days; see every CHL device- ABL90 FLEX PLUS System, safeCLINITUBESK252207 · Radiometer Medicals Aps · Apr 10, 2026
- i-STAT CG4+ cartridge with the i-STAT 1 SystemK244014 · Abbott Point of Care, Inc. · May 2, 2025
- ABL90 FLEX PLUS SystemK240998 · Radiometer Medicals Aps · Dec 13, 2024
- i-STAT CG8+ cartridge with the i-STAT 1 SystemK230285 · Abbott Point of Care, Inc. · Oct 27, 2023
- Stat Profile Prime Plus Analyzer SystemK221900 · Nova Biomedical Corporation · Sep 29, 2023
- i-STAT G3+ cartridge with the i-STAT 1 SystemK223857 · Abbott Point of Care, Inc. · Sep 15, 2023
- GEM Premier 7000 with IQM3K223608 · Instrumentation Laboratory Company · Aug 10, 2023
- EasyStat 300K211559 · Medica Corporation · Jan 27, 2023
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Jan 22, 2016
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1120
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files