ABL90 Flex Plus

Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph. 510(k) K160153, from Radiometer Medical Aps (Broenshoej, DK). Product code CHL, class 2.

As of Nov 17, 2026, K160153 is an FDA 510(k) held by Radiometer Medical Aps, substantially equivalent on Nov 4, 2016 after 287 days, with 5 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Nov 4, 2016
  • Review time: 287 days
  • Recalls: 5
  • Cited as a predicate by: 3 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The ABL90 FLEX PLUS analyzer is an in vitro diagnostic, portable, automated analyser that quantitatively measures, pH, blood gases, electrolytes, glucose, lactate and oximetry in heparinized whole blood, and neonatal bilirubin in heparinized capillary whole blood. The ABL90 FLEX PLUS analyzer is intended for use by trained technologists, nurses, physicians and therapists.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K160153, newest first; each opens FDA's recall record
Mar 21, 2025Z-1659-2025ABL90 FLEX and ABL90 FLEX PLUS analyzers running with Radiometer Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze (the analyzer ceases to operate) during a sample measurement. The analyzer software…Open, Classified
Apr 14, 2021Z-1858-2021Due to an issue with the internal clock of the analyzers with an e3800 PC unit installed.Open, Classified
Apr 14, 2021Z-1857-2021Due to an issue with the internal clock of the analyzers with an e3800 PC unit installed.Open, Classified
Feb 3, 2020Z-1761-2020The firm received reports of occurrences where the barcode reader misinterpreted the contents of a locally printed barcode label used for entering patient ID or accession number into the analyzer in connection with a…Terminated
Nov 21, 2019Z-2138-2020The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has…Terminated

Other CHL Clearances

The newest 510(k)s for the same product code; their median review was 267 days; see every CHL device

Record

Type
Special
Received
Jan 22, 2016
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1120

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files