System, Planning, Radiation Therapy Treatment. 510(k) K160093, from RaySearch Laboratories AB (PUBL) (Stockholm, SE). Product code MUJ, class 2.

As of Nov 20, 2026, K160093 is an FDA 510(k) held by RaySearch Laboratories AB (PUBL), substantially equivalent on Apr 8, 2016 after 80 days, with 7 recalls tied to it in FDA's recall database.

  • Review time: 80 days
  • Recalls: 7
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • RaystationK141860 · Raysearch Laboratories AB · Oct 23, 2014

Cited as a predicate

1 later 510(k) summary names this device as a predicate
  • RayStation 6K170355 · RaySearch Laboratories AB (PUBL) · Mar 2, 2017

Recalls

Recalls FDA ties to K160093, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jan 16, 2025Z-1194-2025Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT…Open
Jan 18, 2017Z-1815-2017An error in NVIDIA GPU (Graphics Processing Unit) card drivers can occur for certain software programs. In a non-standard Citrix environment where the GPU settings are configured with a virtual Citrix graphics board…Terminated
Nov 9, 2016Z-0721-2017Software anomaly; an issue was found with the proton Pencil Beam Scanning (PBS) dose calculation in RayStation 4.0, 4.5, 4.7 and 5.0. For treatment plans with a combination of range shifter, large air gap and beams that…Terminated
Oct 11, 2016Z-0720-2017An error may occur with the display of dose computed on images other than the planning CT (auxiliary CT) when using multiple patient cases in RayStation 5. If a CT image set with the same Frame of Reference as the…Terminated
Sep 1, 2016Z-0038-2017RaySearch became aware of the problem as it was discovered in cooperation with a customer experimenting with the dose calculation on a phantom. The correction concerns two issues found with the dose calculation when…Terminated
Jul 1, 2016Z-2206-2016An issue was found with the evaluation of biological clinical goals in RayStation 3.0, RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7 and RayStation 5. Biological clinical goals for an adapted plan based…Terminated
May 30, 2016Z-0079-2017For a treatment plan consisting of multiple beam sets, the table for ROI plan dose statistics in the report may show the statistics for a beam set dose. The error can only be triggered when using a report template where…Terminated

Other MUJ clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every MUJ device

Record

Type
Traditional
Received
Jan 19, 2016
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files