System, Planning, Radiation Therapy Treatment. 510(k) K141860, from Raysearch Laboratories AB (Stockholm, SE). Product code MUJ, class 2.
As of Nov 20, 2026, K141860 is an FDA 510(k) held by Raysearch Laboratories AB, substantially equivalent on Oct 23, 2014 after 105 days, with 5 recalls tied to it in FDA's recall database.
- Review time: 105 days
- Recalls: 5
- Cited as a predicate by: 1 device
RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs.
- Ray StationK140187 · Raysearch Laboratories AB · May 15, 2014
- RayStationK160093 · RaySearch Laboratories AB (PUBL) · Apr 8, 2016
| Jan 16, 2025 | Z-1195-2025Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT… | Open |
| Feb 10, 2016 | Z-1712-2016A software issue with editing tools that use the left mouse button held down, for drawing in and interacting with the patient views and beams eye views. The views can become unsynchronized with the stored data if… | Terminated |
| May 29, 2015 | Z-2785-2015An error caused by a floating point precision problem tends to cause a problem with the display of isodose lines for dose and dose difference and color tables for dose, dose difference and PET images | Terminated |
| May 27, 2015 | Z-2457-2015Issue with the algorithm for ROI contraction in RayStation 4.7 when non-uniform contraction distances are used. The ROI contraction tool uses six distances as input: right/left, inferior/superior and posterior/anterior.… | Terminated |
| Mar 4, 2015 | Z-1310-2015An issue with photon dose calculation for DMLC (Dynamic MLC) plans for machines where the MLC is positioned above the jaws, e.g. some Elekta linacs. The magnitude of the error depends on the beam model output factor… | Terminated |
- RT Elements (4.6)K261655 · Brainlab SE · Oct 5, 2026
- Virtual Cone (1.0)K260750 · Varian Medical Systems, Inc. · Jul 8, 2026
- RadOncAIK253050 · Informai · Jun 18, 2026
- Multimodality Simulation Workspace (MM Sim) (v1.0.0)K260207 · Philips Medical Systems Nederland B.V. · May 27, 2026
- eXaSkin PlusK252784 · Anatomical Geometry S.L. · May 22, 2026
- ART.1-USK253078 · Seetreat Pty, Ltd. · May 14, 2026
- syngo.via RT Image Suite VC10K252304 · Varian Medical Systems, Inc. · Mar 18, 2026
- Monaco RTP System (6.3)K252002 · Elekta Solutions AB · Feb 19, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jul 10, 2014
- Review panel
- Radiology
- Regulation
- 21 CFR 892.5050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files