Arca
Skip to content
Tools
Tools

2008K2 Hemodialysis Machine

Dialyzer, High Permeability With Or Without Sealed Dialysate System. 510(k) K153449, from Fresenius Medical Care Renal Therapies Group, LLC (Waltham, MA). Product code KDI, class 2.

As of Nov 19, 2026, K153449 is an FDA 510(k) held by Fresenius Medical Care Renal Therapies Group, LLC, substantially equivalent on Apr 26, 2016 after 148 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 148 days
  • Recalls: 2
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The 2008K2 Hemodialysis Machine is indicated for acute and chronic hemodialysis therapy in a healthcare facility. Additional therapy options for patients receiving hemodialysis include: Isolated Ultrafiltration, Sustained Low Efficiency Dialysis (SLED), and low volume hemodialysis (patients weighing • 20 kg and ” 40 kg). This machine accommodates the use of both low flux and high flux dialyzers.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K153449, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jan 29, 2020Z-1529-2020The firm is replacing pressure transducers with an improved design due to an increase in customer complaints regarding lower long-term reliability of the part.Terminated
Dec 14, 2017Z-1278-2018When the recirculation ultrafiltration (UF) Goal is set to a value greater than 200ml in service mode and the user starts a treatment using the SLED (Sustained Low Efficiency Dialysis) program the display will show an…Terminated

Other KDI clearances

The newest 510(k)s for the same product code; their median review was 189 days; see every KDI device

Record

Type
Traditional
Received
Nov 30, 2015
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.5860

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files