Arca
Skip to content
Tools
Tools

2008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T Hemodialysis Machine

Dialyzer, High Permeability With Or Without Sealed Dialysate System. 510(k) K150708, from Fresenius Medical Care Renal Therapies Group, LLC (Waltham, MA). Product code KDI, class 2.

As of Nov 19, 2026, K150708 is an FDA 510(k) held by Fresenius Medical Care Renal Therapies Group, LLC, substantially equivalent on Dec 10, 2015 after 266 days, with 12 recalls tied to it in FDA's recall database.

  • Review time: 266 days
  • Recalls: 12
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

2008KHemodialysis Machines: The 2008K Hemodialysis Machine is indicated for acute and chronic dialysis therapy. 2008K@home Hemodialysis Machine: The 2008K@home Hemodialysis Machine is indicated for acute and chronic dialysis therapy in an acute or chronic facility.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K150708, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Sep 6, 2023Z-0034-2024, Z-0029-2024, Z-0035-2024, Z-0028-2024, Z-0031-2024, Z-0030-2024, Z-0032-2024, Z-0033-2024Potential PCBA leaching from tubing of hemodialysis machines.Open
Dec 21, 2016Z-1029-2017, Z-1026-2017, Z-1027-2017, Z-1028-2017When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machineTerminated

Other KDI clearances

The newest 510(k)s for the same product code; their median review was 189 days; see every KDI device

Record

Type
Traditional
Received
Mar 19, 2015
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.5860

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files