Giraffe OmniBed Carestation CS1
Incubator, Neonatal. 510(k) K152814, from Ohmeda Medical, A Division of Datex-Ohmeda,Inc. A GE Company (Laurel, MD). Product code FMZ, class 2.
As of Nov 18, 2026, K152814 is an FDA 510(k) held by Ohmeda Medical, A Division of Datex-Ohmeda,Inc. A GE Company, substantially equivalent on Mar 17, 2016 after 171 days, with 8 recalls tied to it in FDA's recall database.
- Substantially equivalent: Mar 17, 2016
- Review time: 171 days
- Recalls: 8
- Cited as a predicate by: 1 device
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Giraffe OmniBed Carestation is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user’s demand. It cannot be operated in both modes at the same time.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Giraffe OmnibedK101788 · Ohmeda Medical · Jul 21, 2010
- Ohmeda Medical Giraffe OmnibedK071175 · GE Healthcare · Jul 17, 2007
- Ohmeda Medical Giraffe OmnibedK020543 · Ohmeda Medical · May 9, 2002
- Ohmeda Medical OmnibedK993407 · Ohmeda Medical · Jan 5, 2000
Cited as a Predicate
1 later 510(k) summary names this device as a predicate- Giraffe Incubator Carestation CS1K213553 · Datex Ohmeda, Inc. · Mar 2, 2022
Recalls
Recalls FDA ties to K152814, newest first; each opens FDA's recall record| May 1, 2026 | Z-2477-2026GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners… | Open, Classified |
| May 1, 2026 | Z-2476-2026GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners… | Open, Classified |
| Sep 3, 2024 | Z-0048-2025GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This… | Open, Classified |
| Jun 6, 2023 | Z-0140-2025Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week.… | Completed |
| May 19, 2023 | Z-2100-2023GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe… | Open, Classified |
| Oct 11, 2019 | Z-0539-2020Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall. | Terminated |
| Jul 26, 2019 | Z-0281-2020Fasteners potentially torqued to a value less than the specified value and can potentially cause the canopy to become detached and fall. | Open, Classified |
| Sep 25, 2018 | Z-0479-2019Potential electric leakage affecting the user. | Terminated |
Other FMZ Clearances
The newest 510(k)s for the same product code; their median review was 179 days; see every FMZ device- Giraffe OmniBed Carestation (CS1); Giraffe Incubator Carestation (CS1)K260326 · Datex-Ohmeda, Inc. · Jul 28, 2026
- Babyleo TN500K251619 · Drägerwerk AG & Co. KGaA · Feb 13, 2026
- Giraffe OmniBed Carestation (CS1); Giraffe Incubator Carestation (CS1)K251663 · Datex-Ohmeda, Inc. · Oct 22, 2025
- mOm Essential Incubator (ME1)K243437 · Mom Incubators Limited · Aug 14, 2025
- Giraffe Incubator Carestation CS1K213553 · Datex Ohmeda, Inc. · Mar 2, 2022
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Sep 28, 2015
- Review panel
- General Hospital
- Regulation
- 21 CFR 880.5400
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files