Arca
Skip to content
Tools
Tools

Giraffe Omnibed

Incubator, Neonatal. 510(k) K101788, from Ohmeda Medical (Laurel, MD). Product code FMZ, class 2.

As of Nov 18, 2026, K101788 is an FDA 510(k) held by Ohmeda Medical, substantially equivalent on Jul 21, 2010 after 26 days, with 8 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jul 21, 2010
  • Review time: 26 days
  • Recalls: 8
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user's demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology.

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K101788, newest first; each opens FDA's recall record
May 1, 2026Z-2475-2026GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners…Open, Classified
May 1, 2026Z-2477-2026GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners…Open, Classified
Sep 3, 2024Z-0047-2025GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This…Open, Classified
May 19, 2023Z-2101-2023GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe…Open, Classified
Apr 28, 2020Z-2235-2020Potential for certain Giraffe Incubators and Giraffe OmniBeds to deliver a different environmental oxygen level than what is displayed by the device, if a certain optional component referred to as the Servo Oxygen…Terminated
Apr 2, 2020Z-2111-2020During Giraffe Shuttle charging, the batteries emit low concentrations of gases (outgassing) producing a strong sulfur odor.Terminated
Oct 11, 2019Z-0538-2020Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall.Terminated
Jan 8, 2013Z-1051-2013Neonatal beds and incubator display boards may randomly overwrite stored information and radiant heater and oxygen set points upon power up or recovery from a power failure.Terminated

Other FMZ Clearances

The newest 510(k)s for the same product code; their median review was 179 days; see every FMZ device

Record

Type
Special
Received
Jun 25, 2010
Review panel
General Hospital
Regulation
21 CFR 880.5400

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.