ViperWire Advance Guide Wire, Diamondback 360 Peripheral Orbital Atherectomy System, Stealth 360 Orbital PAD
Catheter, Peripheral, Atherectomy. 510(k) K151260, from Cardiovascular System, Inc. (Saint Paul, MN). Product code MCW, class 2.
As of Nov 19, 2026, K151260 is an FDA 510(k) held by Cardiovascular System, Inc., substantially equivalent on Jun 10, 2015 after 29 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 29 days
- Recalls: 2
- Cited as a predicate by: 1 device
The Stealth 360 Orbital PAD System and the Diamondback 360 Peripheral Orbital Atherectomy System are percutaneous orbital atherectomy systems indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.
- Stealth 360 Orbital Pad SystemK110389 · Cardiovascular Systems, Inc. · Mar 11, 2011
- Stealth 360 Orbital Pad SystemK122987 · Cardiovascular Systems, Inc. · Oct 25, 2012
- Stealth 360 Orbital Pad SystemK131092 · Cardio Systems, Inc. · Jun 24, 2013
- Diamondback 360 Peripheral Orbital Atherectomy SystemK133399 · Cardiovascular Systems Incorporated · Nov 22, 2013
- Diamondback 360 Peripheral Orbital Atherectomy SystemK150732 · Cardiovascular Systems, Inc. · Apr 13, 2015
- ViperWire Advance Guide Wire, Stealth 360 Orbital Peripheral Atherectomy Device (PAD), DIAMONDBACK 360 Peripheral…K180416 · Cardiovascular Systems, Inc. · Apr 17, 2018
| Mar 3, 2017 | Z-1771-2017Cardiovascular Systems, Inc. is recalling one lot of ViperWire Advance with Flextip, model VPR-GW-FT18 because it may contain a 0.014" ViperWire Advance Flextip product instead of the 0.018" ViperWire Advance with… | Terminated |
| Feb 9, 2017 | Z-1877-2017CSI is recalling three lots of VIPERWIRE ADVANCE WITH FLEXTIP PERIPHERAL ATHERECTOMY GUIDEWIRE . It was discovered that 0.018in ViperWire Advance with Flextip product was packaged as a 0.014in ViperWire Advance with… | Terminated |
- Auryon Atherectomy SystemK260244 · Eximo Medical · Apr 10, 2026
- FreedomFlow™ Orbital Circumferential Atherectomy System (H6005/ 4Fr 3-Sphere Configuration)K250723 · Cardio Flow Inc., · Apr 25, 2025
- Turbo-Elite Laser Atherectomy CatheterK250385 · Spectranetics · Mar 13, 2025
- Rotarex Atherectomy SystemK242757 · Bard Peripheral Vascular, Inc. · Jan 30, 2025
- FreedomFlow™ Orbital Circumferential Atherectomy SystemK242947 · Cardio Flow Inc., · Nov 5, 2024
- Auryon Atherectomy Catheter 1.7mm; Auryon Atherectomy Catheter 1.7mm, Hydrophilic CoatingK241553 · Eximo Medical, Ltd. · Jun 27, 2024
- FreedomFlow Orbital Circumferential Atherectomy System (H6004/5Fr 3-Sphere Configuration)K233483 · Cardio Flow Inc., · Feb 15, 2024
- Auryon Atherectomy Catheter 1.5 mm XL, Hydrophilic Coating (EXM-4010-H000), Auryon Atherectomy Catheter 0.9 mm XL,…K233668 · Eximo Medical, Ltd. · Dec 15, 2023
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- May 12, 2015
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.4875
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files