Auryon Atherectomy Catheter 1.7mm and 1 more

Catheter, Peripheral, Atherectomy. 510(k) K241553, from Eximo Medical, Ltd. (Rehovot, IL). Product code MCW, class 2.

As of Nov 19, 2026, K241553 is an FDA 510(k) held by Eximo Medical, Ltd., substantially equivalent on Jun 27, 2024 after 27 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 27 days
  • Recalls: 1

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Auryon™ Atherectomy System and Auryon Atherectomy Catheters with aspiration are indicated for use as atherectomy devices for arterial stenoses, including in-stent restenosis (ISR), and to aspirate thrombus adjacent to stenoses in native and stented infra-inguinal arteries.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K241553, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Nov 25, 2024Z-0770-2025Potential for procedural delays or interruptions during use of the Auryon Atherectomy System due to the Auryon Atherectomy System not advancing to the Activation (Ready) mode.Open

Other MCW clearances

The newest 510(k)s for the same product code; their median review was 73 days; see every MCW device

Record

Type
Special
Received
May 31, 2024
Review panel
Cardiovascular
Regulation
21 CFR 870.4875

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.